表观扩散系数在临床前成像系统上具有高度可重复性:来自一项七中心多厂商研究的证据。

Apparent diffusion coefficient is highly reproducible on preclinical imaging systems: Evidence from a seven-center multivendor study.

作者信息

Doblas Sabrina, Almeida Gilberto S, Blé François-Xavier, Garteiser Philippe, Hoff Benjamin A, McIntyre Dominick J O, Wachsmuth Lydia, Chenevert Thomas L, Faber Cornelius, Griffiths John R, Jacobs Andreas H, Morris David M, O'Connor James P B, Robinson Simon P, Van Beers Bernard E, Waterton John C

机构信息

Laboratory of imaging biomarkers, UMR 1149 - CRI, Inserm, Paris Diderot University, Paris, France.

Division of Radiotherapy and Imaging, Institute of Cancer Research, London, UK.

出版信息

J Magn Reson Imaging. 2015 Dec;42(6):1759-64. doi: 10.1002/jmri.24955. Epub 2015 May 26.

Abstract

PURPOSE

To evaluate between-site agreement of apparent diffusion coefficient (ADC) measurements in preclinical magnetic resonance imaging (MRI) systems.

MATERIALS AND METHODS

A miniaturized thermally stable ice-water phantom was devised. ADC (mean and interquartile range) was measured over several days, on 4.7T, 7T, and 9.4T Bruker, Agilent, and Magnex small-animal MRI systems using a common protocol across seven sites. Day-to-day repeatability was expressed as percent variation of mean ADC between acquisitions. Cross-site reproducibility was expressed as 1.96 × standard deviation of percent deviation of ADC values.

RESULTS

ADC measurements were equivalent across all seven sites with a cross-site ADC reproducibility of 6.3%. Mean day-to-day repeatability of ADC measurements was 2.3%, and no site was identified as presenting different measurements than others (analysis of variance [ANOVA] P = 0.02, post-hoc test n.s.). Between-slice ADC variability was negligible and similar between sites (P = 0.15). Mean within-region-of-interest ADC variability was 5.5%, with one site presenting a significantly greater variation than the others (P = 0.0013).

CONCLUSION

Absolute ADC values in preclinical studies are comparable between sites and equipment, provided standardized protocols are employed.

摘要

目的

评估临床前磁共振成像(MRI)系统中表观扩散系数(ADC)测量的不同站点间的一致性。

材料与方法

设计了一种小型热稳定冰水模体。在4.7T、7T和9.4T的布鲁克、安捷伦和马格内克斯小动物MRI系统上,通过一个通用方案在七个站点上,在几天内测量ADC(平均值和四分位间距)。日常重复性以采集之间平均ADC的百分比变化表示。跨站点再现性以ADC值百分比偏差的1.96倍标准差表示。

结果

所有七个站点的ADC测量值相当,跨站点ADC再现性为6.3%。ADC测量的平均日常重复性为2.3%,未发现有站点的测量值与其他站点不同(方差分析[ANOVA]P = 0.02,事后检验无显著性差异)。层间ADC变异性可忽略不计,且站点间相似(P = 0.15)。感兴趣区内ADC的平均变异性为5.5%,有一个站点的变异性显著大于其他站点(P = 0.0013)。

结论

如果采用标准化方案,临床前研究中的绝对ADC值在不同站点和设备之间具有可比性。

相似文献

1
2
Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice-water phantom.
J Magn Reson Imaging. 2013 May;37(5):1238-46. doi: 10.1002/jmri.23825. Epub 2012 Sep 28.
3
Quality assurance multicenter comparison of different MR scanners for quantitative diffusion-weighted imaging.
J Magn Reson Imaging. 2016 Jan;43(1):213-9. doi: 10.1002/jmri.24956. Epub 2015 May 26.
9
Diffusion tensor imaging of the median nerve at 3.0 T using different MR scanners: agreement of FA and ADC measurements.
Eur J Radiol. 2013 Oct;82(10):e590-6. doi: 10.1016/j.ejrad.2013.05.011. Epub 2013 Jun 12.
10
Apparent diffusion coefficient reproducibility of the pancreas measured at different MR scanners using diffusion-weighted imaging.
J Magn Reson Imaging. 2014 Dec;40(6):1375-81. doi: 10.1002/jmri.24492. Epub 2013 Nov 12.

引用本文的文献

2
Automated quality control of small animal MR neuroimaging data.
Imaging Neurosci (Camb). 2024 Oct 17;2:1-23. doi: 10.1162/imag_a_00317. eCollection 2024 Oct 1.
4
7
Repeatability and reproducibility of ADC measurements: a prospective multicenter whole-body-MRI study.
Eur Radiol. 2021 Jul;31(7):4514-4527. doi: 10.1007/s00330-020-07522-0. Epub 2021 Jan 6.
8
Repeatability and reproducibility of longitudinal relaxation rate in 12 small-animal MRI systems.
Magn Reson Imaging. 2019 Jun;59:121-129. doi: 10.1016/j.mri.2019.03.008. Epub 2019 Mar 12.
9
Recommendations towards standards for quantitative MRI (qMRI) and outstanding needs.
J Magn Reson Imaging. 2019 Jun;49(7):e26-e39. doi: 10.1002/jmri.26598. Epub 2019 Jan 24.
10
Simultaneous characterization of tumor cellularity and the Warburg effect with PET, MRI and hyperpolarized C-MRSI.
Theranostics. 2018 Sep 9;8(17):4765-4780. doi: 10.7150/thno.25162. eCollection 2018.

本文引用的文献

1
Comparison between diffusion-weighted MRI (DW-MRI) at 1.5 and 3 tesla: a phantom study.
J Magn Reson Imaging. 2014 Sep;40(3):682-90. doi: 10.1002/jmri.24397. Epub 2013 Oct 31.
2
The emerging science of quantitative imaging biomarkers terminology and definitions for scientific studies and regulatory submissions.
Stat Methods Med Res. 2015 Feb;24(1):9-26. doi: 10.1177/0962280214537333. Epub 2014 Jun 11.
6
Assessing reproducibility of diffusion-weighted magnetic resonance imaging studies in a murine model of HER2+ breast cancer.
Magn Reson Imaging. 2014 Apr;32(3):245-9. doi: 10.1016/j.mri.2013.10.013. Epub 2013 Dec 14.
7
Apparent diffusion coefficient reproducibility of the pancreas measured at different MR scanners using diffusion-weighted imaging.
J Magn Reson Imaging. 2014 Dec;40(6):1375-81. doi: 10.1002/jmri.24492. Epub 2013 Nov 12.
10
Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice-water phantom.
J Magn Reson Imaging. 2013 May;37(5):1238-46. doi: 10.1002/jmri.23825. Epub 2012 Sep 28.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验