临床实践中针对飞利浦Brilliance大孔径CT的螺旋4D CT螺距管理

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice.

作者信息

Hilgers Guido, Nuver Tonnis, Minken André

机构信息

Radiotherapiegroep | Deventer.

出版信息

J Appl Clin Med Phys. 2015 May 8;16(3):5111. doi: 10.1120/jacmp.v16i3.5111.

Abstract

In external beam radiotherapy treatment planning for patients with thoracic malignancies, respiratory-correlated CT (4D CT) is used to obtain high quality studies in the presence of respiratory motion. When helical 4D CT scans are acquired with a Brilliance CT Big Bore, the pitch must meet two conditions. It must be low enough to avoid motion artifacts, and high enough to cover the entire scan length within 120 s to prevent overheating of the X-ray tube. We developed a nomogram that can be used to obtain a suitable pitch satisfying both requirements. We also assessed the effects on the image quality of a pitch that exceeds the maximum pitch, and of a field of view (FOV) reduction. It was shown that, for AVG and MIP reconstructions, the manufacturer's maximum pitch equation yields an underestimation due to its FOV term.

摘要

在胸部恶性肿瘤患者的外照射放疗治疗计划中,呼吸相关CT(4D CT)用于在存在呼吸运动的情况下获得高质量的研究。当使用Brilliance CT大孔径扫描仪进行螺旋4D CT扫描时,螺距必须满足两个条件。它必须足够低以避免运动伪影,并且足够高以在120秒内覆盖整个扫描长度,以防止X射线管过热。我们开发了一种列线图,可用于获得满足这两个要求的合适螺距。我们还评估了超过最大螺距的螺距以及视野(FOV)减小对图像质量的影响。结果表明,对于AVG和MIP重建,由于其FOV项,制造商的最大螺距公式会产生低估。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fabe/5690122/8dde10cfccd7/ACM2-16-389-g001.jpg

相似文献

1
Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice.
J Appl Clin Med Phys. 2015 May 8;16(3):5111. doi: 10.1120/jacmp.v16i3.5111.
3
Helical mode lung 4D-CT reconstruction using Bayesian model.
Med Image Comput Comput Assist Interv. 2013;16(Pt 3):33-40. doi: 10.1007/978-3-642-40760-4_5.
4
Comparison of helical and cine acquisitions for 4D-CT imaging with multislice CT.
Med Phys. 2005 Feb;32(2):627-34. doi: 10.1118/1.1855013.
5
7
Digital reconstruction of high-quality daily 4D cone-beam CT images using prior knowledge of anatomy and respiratory motion.
Comput Med Imaging Graph. 2015 Mar;40:30-8. doi: 10.1016/j.compmedimag.2014.10.007. Epub 2014 Oct 29.
9
Respiratory correlated cone beam CT.
Med Phys. 2005 Apr;32(4):1176-86. doi: 10.1118/1.1869074.

引用本文的文献

1
A comprehensive quality assurance procedure for 4D CT commissioning and periodic QA.
J Appl Clin Med Phys. 2022 Nov;23(11):e13764. doi: 10.1002/acm2.13764. Epub 2022 Sep 4.
2
Quality assurance of four-dimensional computed tomography in a multicentre trial of stereotactic body radiotherapy of centrally located lung tumours.
Phys Imaging Radiat Oncol. 2018 Dec 7;8:57-62. doi: 10.1016/j.phro.2018.10.003. eCollection 2018 Oct.

本文引用的文献

1
External respiratory motion analysis and statistics for patients and volunteers.
J Appl Clin Med Phys. 2013 Mar 4;14(2):4051. doi: 10.1120/jacmp.v14i2.4051.
2
Overranging at multisection CT: an underestimated source of excess radiation exposure.
Radiographics. 2010 Jul-Aug;30(4):1057-67. doi: 10.1148/rg.304095167.
3
Quality assurance challenges for motion-adaptive radiation therapy: gating, breath holding, and four-dimensional computed tomography.
Int J Radiat Oncol Biol Phys. 2008;71(1 Suppl):S103-7. doi: 10.1016/j.ijrobp.2007.07.2386.
4
The management of respiratory motion in radiation oncology report of AAPM Task Group 76.
Med Phys. 2006 Oct;33(10):3874-900. doi: 10.1118/1.2349696.
5
Acquiring 4D thoracic CT scans using a multislice helical method.
Phys Med Biol. 2004 May 21;49(10):2053-67. doi: 10.1088/0031-9155/49/10/015.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验