The aim of this study was to compare the outcomes of wavefront-optimized sub-Bowman keratomileusis (SBK) performed using the 150 KHz IntraLase femtosecond (IFS) laser (Abbot Medical Optics) with those using the Moria Evo 3 One Use-Plus SBK mechanical microkeratome (Moria SA).
A prospective comparative clinical trial.
Fifty patients had bilateral wavefront-optimized SBK performed for myopia or myopic astigmatism. In 25 patients, the flap was created with the 150 KHz IFS laser (group 1) and in the other 25 with the Moria One Use-Plus SBK mechanical microkeratome (group 2). All other treatment parameters were the same. Results were reported for 1 eye per patient in each group (ie, 25 eyes in each group).
Efficacy, refractive stability, predictability, and safety profiles were comparable between the 2 groups. Although higher order aberrations (HOAs) increased significantly in both groups after laser in situ keratomileusis, intergroup comparison revealed that group 1 was associated with a significantly higher induction of both total HOAs [mean root mean square (RMS), 0.63 ± 0.20 in group 1 vs 0.52 ± 0.14 in group 2; P = 0.031] and spherical aberrations (RMS, 0.37 ± 0.17 in group 1 vs 0.22 ± 0.09 in group 2; P < 0.001) as compared with group 2.
Flaps created using the Moria Evo 3 One Use-Plus SBK mechanical microkeratome were associated with significantly lower induction of total HOAs and spherical aberrations as compared with 150 KHz IFS flaps at the end of 3 months follow-up.
本研究旨在比较使用150千赫兹IntraLase飞秒(IFS)激光(雅培医疗光学公司)进行的波前优化准分子原位角膜磨镶术(SBK)与使用Moria Evo 3一次性使用升级版SBK机械微型角膜刀(Moria SA公司)的手术效果。
一项前瞻性比较临床试验。
50例患者因近视或近视散光接受双侧波前优化SBK手术。25例患者使用150千赫兹IFS激光制作角膜瓣(第1组),另外25例使用Moria一次性使用升级版SBK机械微型角膜刀制作角膜瓣(第2组)。所有其他治疗参数相同。每组报告每位患者一只眼的结果(即每组25只眼)。
两组在疗效、屈光稳定性、可预测性和安全性方面具有可比性。虽然两组在准分子原位角膜磨镶术后高阶像差(HOAs)均显著增加,但组间比较显示,与第2组相比,第1组的总HOAs诱导率[均方根(RMS),第1组为0.63±0.20,第2组为0.52±0.14;P = 0.031]和球差(RMS,第1组为0.37±0.17,第2组为0.22±0.09;P < 0.001)均显著更高。
在3个月随访结束时,与150千赫兹IFS激光制作的角膜瓣相比,使用Moria Evo 3一次性使用升级版SBK机械微型角膜刀制作的角膜瓣在总HOAs和球差诱导方面显著更低。