Abstract

PURPOSE

To assess whether measurements on American College of Radiology (ACR) phantom images performed by magnetic resonance imaging (MRI) technologists as part of a weekly quality control (QC) program could be performed exclusively using an automated system without compromising the integrity of the QC program.

MATERIALS AND METHODS

ACR phantom images are acquired on 15 MRI scanners at a number of ACR-accredited sites to fulfill requirements of a weekly QC program. MRI technologists routinely perform several measurements on these images. Software routines are also used to perform the measurements. A set of geometry measurements made by technologists over a five week period and those made using software routines were compared to reference-standard measurements made by two MRI physicists.

RESULTS

The geometry measurements performed by software routines had a very high positive correlation (0.92) with the reference-standard measurements. Technologist measurements also had a high positive correlation (0.63), although the correlation was less than for the automated measurements. Bland-Altman analysis revealed overall good agreement between the automated and reference-standard measurements, with the 95% limits of agreement being within ±0.62 mm. Agreement between the technologist and the reference-standard measurements was demonstratively poorer, with 95% limits of agreement being ±1.46 mm. Some of the technologist measurements differed from the reference standard by as much as 2 mm.

CONCLUSION

The technologists' geometry measurements may be able to be replaced by automated measurement without compromising the weekly QC program required by the ACR.

摘要

目的

评估磁共振成像(MRI)技术人员在每周质量控制(QC)程序中对美国放射学会(ACR)体模图像进行的测量,是否可以完全使用自动化系统进行,而不影响QC程序的完整性。

材料与方法

在多个获得ACR认可的地点的15台MRI扫描仪上采集ACR体模图像,以满足每周QC程序的要求。MRI技术人员通常对这些图像进行多项测量。也使用软件程序进行测量。将技术人员在五周内进行的一组几何测量结果与使用软件程序进行的测量结果,与两位MRI物理学家进行的参考标准测量结果进行比较。

结果

软件程序进行的几何测量与参考标准测量具有非常高的正相关性(0.92)。技术人员的测量也具有较高的正相关性(0.63),尽管相关性低于自动化测量。Bland-Altman分析显示自动化测量与参考标准测量总体上具有良好的一致性,95%的一致性界限在±0.62毫米以内。技术人员与参考标准测量之间的一致性明显较差,95%的一致性界限为±1.46毫米。一些技术人员的测量结果与参考标准相差多达2毫米。

结论

技术人员的几何测量可以被自动化测量取代,而不影响ACR要求的每周QC程序。