Malyugin Boris E, Shpak Alexander A, Pokrovskiy Dmitry F
From the Cataract & Implant Surgery Department (Malyugin, Pokrovskiy) and Clinical & Functional Diagnostics Department (Shpak), S. Fyodorov Eye Microsurgery Federal State Institution, Moscow, Russia.
From the Cataract & Implant Surgery Department (Malyugin, Pokrovskiy) and Clinical & Functional Diagnostics Department (Shpak), S. Fyodorov Eye Microsurgery Federal State Institution, Moscow, Russia.
J Cataract Refract Surg. 2015 Aug;41(8):1616-22. doi: 10.1016/j.jcrs.2014.12.051.
To use anterior segment optical coherence tomography (AS-OCT) to evaluate the clinical effectiveness of Implantable Collamer Lens posterior chamber phakic intraocular lens (PC pIOL) sizing based on measurement of the distance from the iris pigment end to the iris pigment end.
S. Fyodorov Eye Microsurgery Federal State Institution, Moscow, Russia.
Evaluation of diagnostic test or technology.
Stage 1 was a prospective study. The sulcus-to-sulcus (STS) distance was measured using ultrasound biomicroscopy (UBM) (Vumax 2), and the distance from iris pigment end to iris pigment end was assessed using a proposed AS-OCT algorithm. Part 2 used retrospective data from patients after implantation of a PC pIOL with the size selected according to AS-OCT (Visante) measurements of the distance from iris pigment end to iris pigment end. The PC pIOL vault was measured by AS-OCT, and adverse events were assessed.
Stage 1 comprised 32 eyes of 32 myopic patients (mean age 28.4 years ± 6.3 [SD]; mean spherical equivalent [SE] -13.11 ± 4.28 diopters [D]). Stage 2 comprised 29 eyes of 16 patients (mean age 27.7 ± 4.7 years; mean SE -16.55 ± 3.65 D). The mean STS distance (12.35 ± 0.47 mm) was similar to the mean distance from iris pigment end to iris pigment end distance (examiner 1: 12.36 ± 0.51 mm; examiner 2: 12.37 ± 0.53 mm). The PC pIOL sized using the new AS-OCT algorithm had a mean vault of 0.53 ± 0.18 mm and did not produce adverse events during the 12-month follow-up. In 16 of 29 eyes, the PC pIOL vault was within an optimum interval (0.35 to 0.70 mm).
The new measurement algorithm can be effectively used for PC pIOL sizing.
No author has a financial or proprietary interest in any material or method mentioned.
使用眼前节光学相干断层扫描(AS - OCT),基于测量虹膜色素端到虹膜色素端的距离,评估可植入式胶原晶状体后房型有晶状体眼人工晶状体(PC pIOL)尺寸确定的临床效果。
俄罗斯莫斯科的S. Fyodorov眼显微外科联邦国家机构。
诊断试验或技术评估。
第一阶段为前瞻性研究。使用超声生物显微镜(UBM)(Vumax 2)测量睫状沟到睫状沟(STS)的距离,并使用一种提议的AS - OCT算法评估虹膜色素端到虹膜色素端的距离。第二部分使用植入PC pIOL患者的回顾性数据,PC pIOL的尺寸根据AS - OCT(Visante)测量的虹膜色素端到虹膜色素端的距离来选择。通过AS - OCT测量PC pIOL的拱高,并评估不良事件。
第一阶段包括32例近视患者的32只眼(平均年龄28.4岁±6.3[标准差];平均等效球镜度[SE] - 13.11±4.28屈光度[D])。第二阶段包括16例患者的29只眼(平均年龄27.7±4.7岁;平均SE - 16.55±3.65 D)。平均STS距离(12.35±0.47 mm)与虹膜色素端到虹膜色素端的平均距离相似(检查者1:12.36±0.51 mm;检查者2:12.37±0.53 mm)。使用新的AS - OCT算法确定尺寸的PC pIOL平均拱高为0.53±0.18 mm,在12个月的随访期间未产生不良事件。在29只眼中的16只眼中,PC pIOL的拱高处于最佳区间(0.35至0.70 mm)。
新的测量算法可有效地用于确定PC pIOL的尺寸。
没有作者对文中提及的任何材料或方法拥有财务或专利权益。