Sagiyama Koji, Watanabe Yuji, Kamei Ryotaro, Baba Shingo, Honda Hiroshi
Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
Department of Molecular Imaging and Diagnosis, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
J Magn Reson Imaging. 2016 Apr;43(4):853-8. doi: 10.1002/jmri.25059. Epub 2015 Oct 6.
To compare zoomed diffusion-weighted imaging (z-DWI) with reduced field of view (FOV) by spatially selective radiofrequency pulses and conventional echo planar imaging (EPI) DWI (c-DWI) with regard to registration quality using positron emission tomography / magnetic resonance (PET/MR) in patients with malignant tumors.
Fludeoxyglucose (18(F) ) PET imaging, c-DWI, and z-DWI were conducted simultaneously in 21 patients with known or suspected malignancy using a PET/MR system. A fusion image showing the largest tumor area was generated for analysis. Registration accuracy between PET and DWI was assessed based on the area of maximum overlap and central point displacement of the tumor. EPI factor, echo time (TE), matching area, and displacement were compared between c-DWI and z-DWI by paired t-test. Agreement of apparent diffusion coefficient (ADC) acquired by the two sequences were also assessed with linear regression s and Bland-Altman plot analysis.
Thirty-two lesions were detected on both PET and DWI (mean size 536.3 ± 471.8 mm(2) ). At least one lesion was found in all subjects. In all cases, EPI factor was smaller with z-DWI than c-DWI (43.1 ± 15.6 vs. 62.0 ± 10.0, P < 0.0001), and TE was also shorter for z-DWI (53.6 ± 3.6 msec vs. 65.2 ± 3.6 msec, P < 0.0001). Registration accuracy was better with z-DWI in 30 of 32 lesions (93.8%), and both average matching area and central point displacement were significantly improved (79.8 ± 18.1% vs. 61.8 ± 22.9%, P < 0.0001 and 3.92 ± 2.69 mm vs. 7.51 ± 4.07 mm, P < 0.0001). ADC values calculated with c-DWI and z-DWI showed good agreement.
Zoomed DWI reduces image distortion and provides better registration accuracy with PET images.
比较采用空间选择性射频脉冲的缩小视野(FOV)的放大扩散加权成像(z-DWI)与传统回波平面成像(EPI)扩散加权成像(c-DWI)在恶性肿瘤患者中使用正电子发射断层扫描/磁共振(PET/MR)时的配准质量。
使用PET/MR系统对21例已知或疑似恶性肿瘤患者同时进行氟脱氧葡萄糖(18F)PET成像、c-DWI和z-DWI检查。生成显示最大肿瘤区域的融合图像用于分析。基于肿瘤的最大重叠面积和中心点位移评估PET与DWI之间的配准准确性。通过配对t检验比较c-DWI和z-DWI之间的EPI因子、回波时间(TE)、匹配面积和位移。还通过线性回归和Bland-Altman图分析评估两种序列获取的表观扩散系数(ADC)的一致性。
PET和DWI均检测到32个病变(平均大小536.3±471.8mm2)。所有受试者均发现至少一个病变。在所有病例中,z-DWI的EPI因子均小于c-DWI(43.1±15.6对62.0±10.0,P<0.0001),z-DWI的TE也更短(53.6±3.6毫秒对65.2±3.6毫秒,P<0.0001)。32个病变中的30个(93.8%)z-DWI的配准准确性更好,平均匹配面积和中心点位移均显著改善(79.8±18.1%对61.8±22.9%,P<0.0001;3.92±2.69毫米对7.5l±4.07毫米,P<0.000l)。c-DWI和z-DWI计算的ADC值显示出良好的一致性。
放大扩散加权成像减少了图像失真,并与PET图像提供了更好的配准准确性。