Antonini C J, Brady J M, Levin M P, Garcia W L
J Periodontol. 1977 Jan;48(1):45-8. doi: 10.1902/jop.1977.48.1.45.
Scanning electron microscopy was used to examine factory-sharpened, dulled, and resharpened scalers. The following observations were made: 1. The factory sharpened scalers exhibited metallic extensions from the lateral surface (wire edges). 2. The machine-resharpened instruments resulted in nonfunctional wire edges from the facial surface. Honing produced an exact meeting of facial and lateral surfaces. 3. Hand sharpening, using an Arkansas stone (grit 500) against the lateral surface away from the cutting edge, effected an exact meeting of the facial and lateral surfaces and no evidence of wire edges. 4. From this study an optimum cutting edge for a scaler is characterized as having a smooth, contiguous meeting of the facial and lateral surfaces free of wire edges. If a wire edge is present, it should be functional.
使用扫描电子显微镜检查工厂磨锐、钝后再磨锐的洁治器。观察结果如下:1. 工厂磨锐的洁治器侧面有金属延伸物(丝边)。2. 机器再磨锐的器械在正面产生无功能的丝边。珩磨使正面和侧面精确贴合。3. 使用阿肯色石(粒度500)在远离刃口的侧面进行手工磨锐,可使正面和侧面精确贴合,且无丝边迹象。4. 从本研究可知,洁治器的最佳刃口特征为正面和侧面平滑、连续贴合且无丝边。若有丝边存在,其应具备功能。