Wiant David, Wentworth Stacy, Liu Han, Sintay Benjamin
Cone Health Cancer Center, Greensboro, North Carolina.
Cone Health Cancer Center, Greensboro, North Carolina.
Int J Radiat Oncol Biol Phys. 2015 Nov 15;93(4):901-7. doi: 10.1016/j.ijrobp.2015.06.010. Epub 2015 Oct 19.
Deep inspiration breath hold (DIBH) for left-sided breast cancer has been shown to reduce heart dose. Surface imaging helps to ensure accurate breast positioning, but it does not guarantee a reproducible breath hold (BH) at DIBH treatments. We examine the effects of variable BH positions for DIBH treatments.
Twenty-five patients who underwent free breathing (FB) and DIBH scans were reviewed. Four plans were created for each patient: FB, DIBH, FB_DIBH (the DIBH plans were copied to the FB images and recalculated, and image registration was based on breast tissue), and P_DIBH (a partial BH with the heart shifted midway between the FB and DIBH positions). The FB_DIBH plans give a "worst-case" scenario for surface imaging DIBH, where the breast is aligned by surface imaging but the patient is not holding their breath. Kolmogorov-Smirnov tests were used to compare the dose metrics.
The DIBH plans gave lower heart dose and comparable breast coverage versus FB in all cases. The FB_DIBH plans showed no significant difference versus FB plans for breast coverage, mean heart dose, or maximum heart dose (P≥.10). The mean heart dose differed between FB_DIBH and FB by <2 Gy for all cases, and the maximum heart dose differed by <2 Gy for 21 cases. The P_DIBH plans showed significantly lower mean heart dose than FB (P<.01). The mean heart doses for the P_DIBH plans were <FB for 22 cases, the maximum dose was <FB for 18 cases.
A DIBH plan delivered to a FB patient setup with surface imaging will yield dosimetry similar to that of a plan created and delivered FB. A DIBH plan delivered with even a partial BH can give reduced heart dose compared with FB techniques.
已证明左侧乳腺癌采用深吸气屏气(DIBH)可降低心脏受照剂量。表面成像有助于确保乳房准确定位,但在DIBH治疗中并不能保证可重复的屏气(BH)。我们研究了DIBH治疗中不同BH位置的影响。
回顾了25例行自由呼吸(FB)和DIBH扫描的患者。为每位患者制定了四个计划:FB、DIBH、FB_DIBH(将DIBH计划复制到FB图像上并重新计算,图像配准基于乳腺组织)和P_DIBH(一种部分屏气,心脏位于FB和DIBH位置之间的中间位置)。FB_DIBH计划给出了表面成像DIBH的“最坏情况”,即乳房通过表面成像对齐,但患者未屏气。采用柯尔莫哥洛夫-斯米尔诺夫检验比较剂量指标。
在所有情况下,DIBH计划的心脏受照剂量均低于FB,且乳腺覆盖情况相当。FB_DIBH计划在乳腺覆盖、平均心脏剂量或最大心脏剂量方面与FB计划相比无显著差异(P≥0.10)。所有病例中,FB_DIBH与FB的平均心脏剂量差异<2 Gy,21例的最大心脏剂量差异<2 Gy。P_DIBH计划的平均心脏剂量显著低于FB(P<0.01)。22例P_DIBH计划的平均心脏剂量<FB,18例的最大剂量<FB。
在表面成像的FB患者设置下实施的DIBH计划所产生的剂量学结果与在FB状态下创建和实施的计划相似。与FB技术相比,即使采用部分屏气实施的DIBH计划也可降低心脏受照剂量。