成像类型对治疗计划中勾画过程的影响。

Influence of the type of imaging on the delineation process during the treatment planning.

作者信息

Jackowiak Weronika, Bąk Bartosz, Kowalik Anna, Ryczkowski Adam, Skórska Małgorzata, Paszek-Widzińska Małgorzata

机构信息

Radiotherapy Department I, Greater Poland Cancer Centre, Poznan, Poland.

Radiotherapy Department II, Greater Poland Cancer Centre, Poznan, Poland.

出版信息

Rep Pract Oncol Radiother. 2015 Sep-Oct;20(5):351-7. doi: 10.1016/j.rpor.2015.05.004. Epub 2015 Jun 18.

Abstract

AIM

The aim of this study was to compare the intra- and interobserver contouring variability for structures with density of organ at risk in two types of tomography: kilovoltage computed tomography (KVCT) versus megavoltage computed tomography (MVCT). The intra- and interobserver differences were examined on both types of tomography for structures which simulate human tissue or organs.

MATERIALS AND METHODS

Six structures with density of the liver, bone, trachea, lung, soft tissue and muscle were created and used. For the measurements, the special water phantom with all structures was designed. To evaluate interobserver variability, five observers delineated the structures in both types of computed tomography (CT).

RESULTS

Intraobserver variability was in the range of 1-14% and was the largest for the liver. The observers segmented larger volumes on MVCT compared with KVCT for the trachea (79.56 ccm vs.74.91 ccm), lung (87.61 vs. 82.50), soft tissue (154.24 vs. 145.47) and muscle (164.01 vs. 157.89). For the liver (98.13 vs. 99.38) and bone (51.86 vs. 67.97), the volume on MVCT was smaller than KVCT. The statistically significant differences between observers were observed for structures with density of the liver, bone and soft tissue on KVCT and for the liver, lung and soft tissue on MVCT. For the structures with density of the trachea and muscles, there were no significant differences for both types of tomography.

CONCLUSIONS

During the contouring process the interobserver and intraobserver contouring uncertainty was larger on MVCT, especially for structures with HU near 80, compared with KVCT.

摘要

目的

本研究旨在比较在两种类型的断层扫描中,即千伏计算机断层扫描(KVCT)与兆伏计算机断层扫描(MVCT),观察者内和观察者间对具有危及器官密度的结构进行轮廓勾画的变异性。在这两种类型的断层扫描中,对模拟人体组织或器官的结构进行观察者内和观察者间差异的检查。

材料与方法

创建并使用了六种具有肝脏、骨骼、气管、肺、软组织和肌肉密度的结构。为进行测量,设计了包含所有结构的特殊水模体。为评估观察者间变异性,五名观察者在两种类型的计算机断层扫描(CT)中对结构进行勾画。

结果

观察者内变异性在1% - 14%范围内,肝脏的变异性最大。与KVCT相比,观察者在MVCT上对气管(79.56立方厘米对74.91立方厘米)、肺(87.61对82.50)、软组织(154.24对145.47)和肌肉(164.01对157.89)勾画的体积更大。对于肝脏(98.13对99.38)和骨骼(51.86对67.97),MVCT上的体积小于KVCT。在KVCT上,观察到肝脏、骨骼和软组织密度的结构以及MVCT上肝脏、肺和软组织密度的结构在观察者之间存在统计学显著差异。对于气管和肌肉密度的结构,两种类型的断层扫描均无显著差异。

结论

与KVCT相比,在轮廓勾画过程中,MVCT上观察者间和观察者内的轮廓勾画不确定性更大,尤其是对于HU接近80的结构。

相似文献

1
Influence of the type of imaging on the delineation process during the treatment planning.
Rep Pract Oncol Radiother. 2015 Sep-Oct;20(5):351-7. doi: 10.1016/j.rpor.2015.05.004. Epub 2015 Jun 18.
7
Real patient data based cross verification of kilovoltage and megavoltage CT calibration for proton therapy.
Phys Med. 2016 Feb;32(2):343-52. doi: 10.1016/j.ejmp.2016.01.477. Epub 2016 Feb 26.

引用本文的文献

1
MR-CT image fusion method of intracranial tumors based on Res2Net.
BMC Med Imaging. 2024 Jul 8;24(1):169. doi: 10.1186/s12880-024-01329-x.
3
Autosegmentation of the rectum on megavoltage image guidance scans.
Biomed Phys Eng Express. 2019 Feb;5(2):025006. doi: 10.1088/2057-1976/aaf1db. Epub 2019 Jan 10.
4
Influence of the contrast agents on treatment planning dose calculations of prostate and rectal cancers.
Rep Pract Oncol Radiother. 2016 Sep-Oct;21(5):441-6. doi: 10.1016/j.rpor.2016.04.004. Epub 2016 May 6.

本文引用的文献

1
Functional imaging in radiation therapy planning for head and neck cancer.
Rep Pract Oncol Radiother. 2013 Nov 9;18(6):376-82. doi: 10.1016/j.rpor.2013.10.007.
2
The importance of accurate treatment planning, delivery, and dose verification.
Rep Pract Oncol Radiother. 2012 Mar 6;17(2):63-5. doi: 10.1016/j.rpor.2012.02.001. eCollection 2012.
3
Tomotherapy archive structure and new software tool for loading and advanced analysis of data contained in it.
Rep Pract Oncol Radiother. 2011 Feb 23;16(2):58-64. doi: 10.1016/j.rpor.2011.01.004. eCollection 2011.
4
Impact of the spinal cord position uncertainty on the dose received during head and neck helical tomotherapy.
J Med Imaging Radiat Oncol. 2013 Aug;57(4):503-11. doi: 10.1111/1754-9485.12056. Epub 2013 Apr 1.
5
Doses in organs at risk during head and neck radiotherapy using IMRT and 3D-CRT.
Radiol Oncol. 2012 Dec;46(4):328-36. doi: 10.2478/v10019-012-0050-y. Epub 2012 Nov 9.
6
An investigation into factors affecting electron density calibration for a megavoltage cone-beam CT system.
J Appl Clin Med Phys. 2012 Sep 6;13(5):3271. doi: 10.1120/jacmp.v13i5.3271.
8
3D Variation in delineation of head and neck organs at risk.
Radiat Oncol. 2012 Mar 13;7:32. doi: 10.1186/1748-717X-7-32.
9
Evaluation of tomotherapy MVCT image enhancement program for tumor volume delineation.
J Appl Clin Med Phys. 2011 Apr 12;12(3):3505. doi: 10.1120/jacmp.v12i3.3505.
10
Dual-energy (spectral) CT: applications in abdominal imaging.
Radiographics. 2011 Jul-Aug;31(4):1031-46; discussion 1047-50. doi: 10.1148/rg.314105159.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验