Beyer Thomas, Lassen Martin L, Boellaard Ronald, Delso Gaspar, Yaqub Maqsood, Sattler Bernhard, Quick Harald H
Center for Medical Physics and Biomedical Engineering, General Hospital Vienna, Medical University of Vienna, Waehringer Guertel 18-20/4L, 1090, Vienna, Austria.
Department of Radiology and Nuclear Medicine, VU University Medical Center, Amsterdam, The Netherlands.
MAGMA. 2016 Feb;29(1):75-87. doi: 10.1007/s10334-015-0505-4. Epub 2016 Jan 6.
We assess inter- and intra-subject variability of magnetic resonance (MR)-based attenuation maps (MRμMaps) of human subjects for state-of-the-art positron emission tomography (PET)/MR imaging systems.
Four healthy male subjects underwent repeated MR imaging with a Siemens Biograph mMR, Philips Ingenuity TF and GE SIGNA PET/MR system using product-specific MR sequences and image processing algorithms for generating MRμMaps. Total lung volumes and mean attenuation values in nine thoracic reference regions were calculated. Linear regression was used for comparing lung volumes on MRμMaps. Intra- and inter-system variability was investigated using a mixed effects model.
Intra-system variability was seen for the lung volume of some subjects, (p = 0.29). Mean attenuation values across subjects were significantly different (p < 0.001) due to different segmentations of the trachea. Differences in the attenuation values caused noticeable intra-individual and inter-system differences that translated into a subsequent bias of the corrected PET activity values, as verified by independent simulations.
Significant differences of MRμMaps generated for the same subjects but different PET/MR systems resulted in differences in attenuation correction factors, particularly in the thorax. These differences currently limit the quantitative use of PET/MR in multi-center imaging studies.
我们评估了用于最先进的正电子发射断层扫描(PET)/磁共振成像(MR)系统的人体受试者基于磁共振(MR)的衰减图(MRμMaps)的受试者间和受试者内变异性。
四名健康男性受试者使用西门子Biograph mMR、飞利浦Ingenuity TF和GE SIGNA PET/MR系统,采用特定产品的MR序列和图像处理算法进行重复MR成像,以生成MRμMaps。计算九个胸部参考区域的总肺容积和平均衰减值。采用线性回归比较MRμMaps上的肺容积。使用混合效应模型研究系统内和系统间的变异性。
部分受试者的肺容积存在系统内变异性(p = 0.29)。由于气管分割不同,受试者间的平均衰减值存在显著差异(p < 0.001)。衰减值的差异导致了明显的个体内和系统间差异,进而转化为校正后的PET活性值的偏差,独立模拟验证了这一点。
为同一受试者但不同PET/MR系统生成的MRμMaps存在显著差异,导致衰减校正因子存在差异,尤其是在胸部。这些差异目前限制了PET/MR在多中心成像研究中的定量应用。