在70厘米宽孔径和60厘米传统孔径的3.0-T磁共振成像仪上使用容积加速Flex序列进行肝脏采集。

Liver acquisition with volume acceleration flex on 70-cm wide-bore and 60-cm conventional-bore 3.0-T MRI.

作者信息

Saito Shigeyoshi, Tanaka Keiko, Hashido Takashi

机构信息

Department of Medical Physics and Engineering, Division of Medical Technology and Science, Course of Health Science, Graduate School of Medicine, Osaka University, 1-7 Yamadaoka, Suita, Osaka, 565-0871, Japan.

Division of Radiology, Department of Medical Technology, Osaka University Hospital, 2-15 Yamadaoka, Suita City, Osaka, 565-0871, Japan.

出版信息

Radiol Phys Technol. 2016 Jul;9(2):154-60. doi: 10.1007/s12194-015-0344-z. Epub 2016 Jan 6.

Abstract

This study aimed to compare the uniformity of fat suppression and image quality between liver acquisition with volume acceleration flex (LAVA-Flex) and LAVA on 60-cm conventional-bore and 70-cm wide-bore 3.0-T magnetic resonance imaging (MRI). The uniformity of fat suppression by LAVA-Flex and LAVA was assessed as the efficiency of suppression of superficial fat at the levels of the liver dome, porta, and renal hilum. Percentage standard deviation (%SD) was calculated using the following equation: %SD (%) = 100 × SD of the regions of interest (ROIs)/mean value of the signal intensity (SI) in the ROIs. Signal-to-noise ratio (SNR) and contrast ratio (CR) were calculated. In the LAVA sequence, the %SD in all slices on wide-bore 3.0-T MRI was significantly higher than that on conventional-bore 3.0-T MRI (P < 0.01). However, there was no significant difference in fat signal uniformity between the conventional and wide-bore scanners when LAVA-Flex was used. In the liver, there were no significant differences in SNR between the two sequences. However, the SNR in the pancreas was lower for the wide-bore scanner than for the conventional-bore scanner for both sequences (P < 0.05). There were no significant differences in CR for the liver and fat between LAVA-Flex and LAVA in both scanners. The CR in the LAVA-Flex images obtained by wide-bore MRI was significantly higher than that in the LAVA-Flex images recorded by conventional-bore MRI (P < 0.001). LAVA-Flex offers more homogenous fat suppression in the upper abdomen than LAVA for both conventional and wide-bore 3.0-T MRI.

摘要

本研究旨在比较容积加速flex序列肝脏采集(LAVA-Flex)与LAVA序列在60厘米常规孔径和70厘米大孔径3.0-T磁共振成像(MRI)上脂肪抑制的均匀性和图像质量。通过LAVA-Flex和LAVA进行脂肪抑制的均匀性评估为肝脏穹顶、肝门和肾门水平浅表脂肪的抑制效率。百分比标准差(%SD)使用以下公式计算:%SD(%)=100×感兴趣区域(ROI)的标准差/ROI中信号强度(SI)的平均值。计算信噪比(SNR)和对比率(CR)。在LAVA序列中,大孔径3.0-T MRI上所有切片的%SD显著高于常规孔径3.0-T MRI(P<0.01)。然而,使用LAVA-Flex时,常规扫描仪和大孔径扫描仪之间的脂肪信号均匀性无显著差异。在肝脏中,两个序列之间的SNR无显著差异。然而,对于两个序列,大孔径扫描仪胰腺中的SNR均低于常规孔径扫描仪(P<0.05)。两种扫描仪中,LAVA-Flex和LAVA在肝脏和脂肪的CR方面均无显著差异。大孔径MRI获得的LAVA-Flex图像中的CR显著高于常规孔径MRI记录的LAVA-Flex图像中的CR(P<0.001)。对于常规和大孔径3.0-T MRI,LAVA-Flex在上腹部比LAVA提供更均匀的脂肪抑制。