The purpose of this study was to retrospectively evaluate the results from a Helical TomoTherapy Hi-Art treatment system relating to quality controls based on daily static and dynamic output checks using statistical process control methods. Individual value X-charts, exponentially weighted moving average charts, and process capability and acceptability indices were used to monitor the treatment system performance. Daily output values measured from January 2014 to January 2015 were considered. The results obtained showed that, although the process was in control, there was an out-of-control situation in the principal maintenance intervention for the treatment system. In particular, process capability indices showed a decreasing percentage of points in control which was, however, acceptable according to AAPM TG148 guidelines. Our findings underline the importance of restricting the acceptable range of daily output checks and suggest a future line of investigation for a detailed process control of daily output checks for the Helical TomoTherapy Hi-Art treatment system.
本研究的目的是回顾性评估螺旋断层放射治疗Hi-Art治疗系统基于使用统计过程控制方法的每日静态和动态输出检查的质量控制结果。使用个体值X控制图、指数加权移动平均图以及过程能力和可接受性指数来监测治疗系统性能。考虑了2014年1月至2015年1月测量的每日输出值。获得的结果表明,尽管过程处于控制之中,但治疗系统的主要维护干预存在失控情况。特别是,过程能力指数显示处于控制状态的点的百分比在下降,不过根据美国医学物理学家协会(AAPM)TG148指南这是可接受的。我们的研究结果强调了限制每日输出检查可接受范围的重要性,并为螺旋断层放射治疗Hi-Art治疗系统每日输出检查的详细过程控制提出了未来的研究方向。