Farrer Alexis I, Almquist Scott, Dillon Christopher R, Neumayer Leigh A, Parker Dennis L, Christensen Douglas A, Payne Allison
Department of Bioengineering, University of Utah, 36 South Wasatch Drive, Room 3100, Salt Lake City, Utah 84112.
Med Phys. 2016 Mar;43(3):1374-84. doi: 10.1118/1.4941013.
This simulation study evaluates the effects of phase aberration in breast MR-guided focused ultrasound (MRgFUS) ablation treatments performed with a phased-array transducer positioned laterally to the breast. A quantification of these effects in terms of thermal dose delivery and the potential benefits of phase correction is demonstrated in four heterogeneous breast numerical models.
To evaluate the effects of varying breast tissue properties on the quality of the focus, four female volunteers with confirmed benign fibroadenomas were imaged using 3T MRI. These images were segmented into numerical models with six tissue types, with each tissue type assigned standard acoustic properties from the literature. Simulations for a single-plane 16-point raster-scan treatment trajectory centered in a fibroadenoma in each modeled breast were performed for a breast-specific MRgFUS system. At each of the 16 points, pressure patterns both with and without applying a phase correction technique were determined with the hybrid-angular spectrum method. Corrected phase patterns were obtained using a simulation-based phase aberration correction technique to adjust each element's transmit phase to obtain maximized constructive interference at the desired focus. Thermal simulations were performed for both the corrected and uncorrected pressure patterns using a finite-difference implementation of the Pennes bioheat equation. The effect of phase correction was evaluated through comparison of thermal dose accumulation both within and outside a defined treatment volume. Treatment results using corrected and uncorrected phase aberration simulations were compared by evaluating the power required to achieve a 20 °C temperature rise at the first treatment location. The extent of the volumes that received a minimum thermal dose of 240 CEM at 43 °C inside the intended treatment volume as well as the volume in the remaining breast tissues was also evaluated in the form of a dose volume ratio (DVR), a DVR percent change between corrected and uncorrected phases, and an additional metric that measured phase spread.
With phase aberration correction applied, there was an improvement in the focus for all breast anatomies as quantified by a reduction in power required (13%-102%) to reach 20 °C when compared to uncorrected simulations. Also, the DVR percent change increased by 5%-77% in seven out of eight cases, indicating an improvement to the treatment as measured by a reduction in thermal dose deposited to the nontreatment tissues. Breast compositions with a higher degree of heterogeneity along the ultrasound beam path showed greater reductions in thermal dose delivered outside of the treatment volume with correction applied than beam trajectories that propagated through more homogeneous breast compositions. An increasing linear trend was observed between the DVR percent change and the phase-spread metric (R(2) = 0.68).
These results indicate that performing phase aberration correction for breast MRgFUS treatments is beneficial for the small-aperture transducer (14.4 × 9.8 cm) evaluated in this work. While all breast anatomies could benefit from phase aberration correction, greater benefits are observed in more heterogeneous anatomies.
本模拟研究评估了在使用侧向放置于乳房的相控阵换能器进行的乳腺磁共振引导聚焦超声(MRgFUS)消融治疗中,相位畸变的影响。在四个异质性乳腺数值模型中,展示了根据热剂量传递对这些影响的量化以及相位校正的潜在益处。
为了评估不同乳腺组织特性对焦点质量的影响,对四名确诊为良性纤维腺瘤的女性志愿者进行了3T磁共振成像。这些图像被分割为具有六种组织类型的数值模型,每种组织类型都被赋予了文献中的标准声学特性。针对每个模拟乳腺中位于纤维腺瘤中心的单平面16点光栅扫描治疗轨迹,使用特定于乳腺的MRgFUS系统进行模拟。在这16个点中的每一个点,使用混合角谱方法确定了应用和未应用相位校正技术时的压力模式。使用基于模拟的相位畸变校正技术获得校正后的相位模式,以调整每个元件的发射相位,从而在期望焦点处获得最大化的相长干涉。使用Pennes生物热方程的有限差分实现,对校正和未校正的压力模式进行热模拟。通过比较在定义的治疗体积内外的热剂量积累来评估相位校正的效果。通过评估在第一个治疗位置实现20°C温度升高所需的功率,比较了使用校正和未校正相位畸变模拟的治疗结果。还以剂量体积比(DVR)、校正和未校正相位之间的DVR百分比变化以及测量相位扩散的另一个指标的形式,评估了在预期治疗体积内接受43°C下最小热剂量240 CEM的体积范围以及其余乳腺组织中的体积。
应用相位畸变校正后,与未校正模拟相比,所有乳腺解剖结构的焦点都有改善,这通过达到20°C所需功率的降低(13%-102%)来量化。此外,在八分之七的病例中,DVR百分比变化增加了5%-77%,这表明通过减少沉积到非治疗组织的热剂量来衡量,治疗得到了改善。与传播通过更均匀乳腺成分的波束轨迹相比,沿着超声束路径具有更高异质性程度的乳腺成分在应用校正后,治疗体积外传递的热剂量减少得更多。在DVR百分比变化和相位扩散指标之间观察到增加的线性趋势(R(2)=0.68)。
这些结果表明,对乳腺MRgFUS治疗进行相位畸变校正对于本研究中评估的小孔径换能器(14.4×9.8 cm)是有益的。虽然所有乳腺解剖结构都可以从相位畸变校正中受益,但在更异质的解剖结构中观察到更大的益处。