使用倾斜凝胶垫优化用于多普勒超声流速测量的射束流角度。

Optimization of beam-flow angles for Doppler ultrasound flow velocity measurements using slanted gel pads.

作者信息

Park Michael Yong, Jung Seung Eun, Choi Joon-Il, Byun Jae Young

机构信息

Department of Radiology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-Daero, Seocho-gu, Seoul, 06591 Korea.

出版信息

Springerplus. 2016 Mar 15;5:328. doi: 10.1186/s40064-016-1977-z. eCollection 2016.

Abstract

The aim of this study was to assess whether slanted gel pads can be used to optimize beam-flow angles and flow velocity measurements for Doppler ultrasound. The right carotid artery of a single healthy female volunteer was measured alternatively five times without and with an 18° angled slanted gel pad between the ultrasound transducer and skin by 13 radiologists. Beam-flow angles and peak systolic flow velocities (PSV) were measured along with assessment of spectral broadening. Beam-flow angles (P = 0.001) and PSV (P = 0.001) measurements showed a significant decrease when using slanted gel pads. The mean (±SD) beam-flow angles without and with the use of slanted gel pads were 66.7 (±4.2) and 56.1 (±5.8) degrees, respectively. The mean (±SD) PSVs without and with the use of slanted gel pads were 92.0 (±17.4) and 76.9 (±10.9) cm/s, respectively. There was a noticeable decrease in spectral broadening when using slanted gel pads. There was a significant linear correlation between beam-flow angle and peak systolic velocity. Coefficients of variation for peak systolic velocity without and with the use of gel pads were 18.9 and 14.2 %, respectively. These results demonstrate that slanted gel pads decrease beam-flow angles and overestimation of Doppler flow velocity measurements while potentially increasing the reliability of measurements.

摘要

本研究的目的是评估倾斜凝胶垫是否可用于优化多普勒超声的声束-血流角度和流速测量。13名放射科医生对一名健康女性志愿者的右侧颈动脉进行了测量,在超声换能器与皮肤之间不使用和使用18°倾斜的凝胶垫的情况下交替测量了5次。测量了声束-血流角度和收缩期峰值流速(PSV),并评估了频谱展宽。使用倾斜凝胶垫时,声束-血流角度(P = 0.001)和PSV(P = 0.001)测量值显著降低。不使用和使用倾斜凝胶垫时的平均(±标准差)声束-血流角度分别为66.7(±4.2)度和56.1(±5.8)度。不使用和使用倾斜凝胶垫时的平均(±标准差)PSV分别为92.0(±17.4)cm/s和76.9(±10.9)cm/s。使用倾斜凝胶垫时频谱展宽明显降低。声束-血流角度与收缩期峰值速度之间存在显著的线性相关性。不使用和使用凝胶垫时收缩期峰值速度的变异系数分别为18.9%和14.2%。这些结果表明,倾斜凝胶垫可减小声束-血流角度,减少多普勒流速测量的高估,同时可能提高测量的可靠性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/28ac/4791444/9e92478186c2/40064_2016_1977_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验