新生儿模型中不同正压通气设备的比较。

Comparison of positive pressure ventilation devices in a newborn manikin.

作者信息

Solevåg Anne Lee, Haemmerle Enrico, van Os Sylvia, Bach Katinka P, Cheung Po-Yin, Schmölzer Georg M

机构信息

a Centre for the Studies of Asphyxia and Resuscitation, Neonatal Research Unit, Royal Alexandra Hospital , Edmonton , Canada.

b Department of Pediatrics , University of Alberta , Edmonton , Canada.

出版信息

J Matern Fetal Neonatal Med. 2017 Mar;30(5):595-599. doi: 10.1080/14767058.2016.1180360. Epub 2016 May 10.

Abstract

OBJECTIVE

To compare tidal volume (V) delivery and ventilation rate between devices for positive pressure ventilation (PPV) during newborn resuscitation.

METHODS

Neonatal resuscitation program providers (n = 25) delivered PPV to a newborn manikin in a randomized order with: a self-inflating bag (SIB), a disposable T-piece, a non-disposable T-piece, a stand-alone infant resuscitation system T-piece and the volume-controlled prototype Next Step device (KM Medical). All T-pieces used a peak inflation pressure of 20cmHO and a 5cmHO positive end-expiratory pressure (PEEP). The SIB neither had a PEEP valve nor manometer. The Next Step had a 5cmHO PEEP valve. The participants aimed to deliver a 5 mL/kg V (rate 40-60 min) for 1 min with each device and each of three compliances (0.5, 1.0 and 2.0 mL/cmHO). V and ventilation rate were compared between devices and compliance levels (ANOVA) Results: All devices, except the Next Step delivered a 4-5 mL/kg V at the low compliance, but three- to four-fold that of the target at the higher compliance levels. The Next Step delivered a V close to target at all compliance levels. The ventilation rate was within 40-60 min with all devices and compliance levels.

CONCLUSIONS

Routinely used ventilation devices for newborn resuscitation can triple intended V and requires further investigation.

摘要

目的

比较新生儿复苏期间用于正压通气(PPV)的设备之间的潮气量(V)输送和通气频率。

方法

新生儿复苏项目提供者(n = 25)以随机顺序对新生儿人体模型进行PPV,使用的设备有:自膨胀气囊(SIB)、一次性T形管、非一次性T形管、独立婴儿复苏系统T形管以及容量控制原型Next Step设备(KM Medical)。所有T形管的峰值充气压力均为20cmH₂O,呼气末正压(PEEP)为5cmH₂O。SIB既没有PEEP阀也没有压力计。Next Step有一个5cmH₂O的PEEP阀。参与者旨在使用每种设备以及三种顺应性(0.5、1.0和2.0mL/cmH₂O)中的每一种,在1分钟内输送5mL/kg的V(频率为40 - 60次/分钟)。比较不同设备和顺应性水平之间的V和通气频率(方差分析)。结果:除Next Step外,所有设备在低顺应性时输送的V为4 - 5mL/kg,但在较高顺应性水平时是目标值的三到四倍。Next Step在所有顺应性水平下输送的V均接近目标值。所有设备和顺应性水平下的通气频率均在40 - 60次/分钟范围内。

结论

常规用于新生儿复苏的通气设备输送的V可能是预期值的三倍,需要进一步研究。