Solevåg Anne Lee, Haemmerle Enrico, van Os Sylvia, Bach Katinka P, Cheung Po-Yin, Schmölzer Georg M
a Centre for the Studies of Asphyxia and Resuscitation, Neonatal Research Unit, Royal Alexandra Hospital , Edmonton , Canada.
b Department of Pediatrics , University of Alberta , Edmonton , Canada.
J Matern Fetal Neonatal Med. 2017 Mar;30(5):595-599. doi: 10.1080/14767058.2016.1180360. Epub 2016 May 10.
To compare tidal volume (V) delivery and ventilation rate between devices for positive pressure ventilation (PPV) during newborn resuscitation.
Neonatal resuscitation program providers (n = 25) delivered PPV to a newborn manikin in a randomized order with: a self-inflating bag (SIB), a disposable T-piece, a non-disposable T-piece, a stand-alone infant resuscitation system T-piece and the volume-controlled prototype Next Step device (KM Medical). All T-pieces used a peak inflation pressure of 20cmHO and a 5cmHO positive end-expiratory pressure (PEEP). The SIB neither had a PEEP valve nor manometer. The Next Step had a 5cmHO PEEP valve. The participants aimed to deliver a 5 mL/kg V (rate 40-60 min) for 1 min with each device and each of three compliances (0.5, 1.0 and 2.0 mL/cmHO). V and ventilation rate were compared between devices and compliance levels (ANOVA) Results: All devices, except the Next Step delivered a 4-5 mL/kg V at the low compliance, but three- to four-fold that of the target at the higher compliance levels. The Next Step delivered a V close to target at all compliance levels. The ventilation rate was within 40-60 min with all devices and compliance levels.
Routinely used ventilation devices for newborn resuscitation can triple intended V and requires further investigation.
比较新生儿复苏期间用于正压通气(PPV)的设备之间的潮气量(V)输送和通气频率。
新生儿复苏项目提供者(n = 25)以随机顺序对新生儿人体模型进行PPV,使用的设备有:自膨胀气囊(SIB)、一次性T形管、非一次性T形管、独立婴儿复苏系统T形管以及容量控制原型Next Step设备(KM Medical)。所有T形管的峰值充气压力均为20cmH₂O,呼气末正压(PEEP)为5cmH₂O。SIB既没有PEEP阀也没有压力计。Next Step有一个5cmH₂O的PEEP阀。参与者旨在使用每种设备以及三种顺应性(0.5、1.0和2.0mL/cmH₂O)中的每一种,在1分钟内输送5mL/kg的V(频率为40 - 60次/分钟)。比较不同设备和顺应性水平之间的V和通气频率(方差分析)。结果:除Next Step外,所有设备在低顺应性时输送的V为4 - 5mL/kg,但在较高顺应性水平时是目标值的三到四倍。Next Step在所有顺应性水平下输送的V均接近目标值。所有设备和顺应性水平下的通气频率均在40 - 60次/分钟范围内。
常规用于新生儿复苏的通气设备输送的V可能是预期值的三倍,需要进一步研究。