Suppr超能文献

两种常用自动诊室血压测量设备在重度肥胖人群中的可比性。

Comparability of two commonly used automated office blood pressure devices in the severely obese.

作者信息

Padwal Raj S, Majumdar Sumit R

机构信息

aDepartment of Medicine, University of Alberta bAlberta Diabetes Institute cMazankowski Heart Institute, Edmonton, Alberta, Canada.

出版信息

Blood Press Monit. 2016 Oct;21(5):313-5. doi: 10.1097/MBP.0000000000000196.

Abstract

The accuracy of automated office blood pressure (BP) devices in severely obese individuals is unknown. In an exploratory study, we compared standardized BP measurements using a BpTRU device with those taken using a WatchBP Office monitor. Paired t-tests, Bland-Altman plots, and Pearson's correlations were used to compare devices. A total of 62 patients were analyzed: their mean age was 39.5±9.3 years and their mean BMI was 45.1±5.7 kg/m. Fifty-five patients (89%) were women and 21 (34%) had a diagnosis of hypertension. The mean BpTRU values were 117.2±12.7/76.7±9.7 mmHg and the mean WatchBP measurements were 126.6±13.8/79.7±10.9 mmHg [difference of -9.4±11.6 for systolic (P<0.0001) and -3.0±7.7 mmHg for diastolic (P=0.003)]. A Bland-Altman plot for systolic BP showed considerable variability. Only four (6.5%) patients had hypertension (>140/90 mmHg) according to BpTRU versus 15 (24%) according to WatchBP. In conclusion, the mean BP measurements taken using two commonly used automated office devices differed markedly and by a clinically important magnitude. These data indicate the need for formal validation of these devices specifically in severely obese individuals.

摘要

在严重肥胖个体中,自动办公室血压测量设备的准确性尚不清楚。在一项探索性研究中,我们将使用BpTRU设备进行的标准化血压测量与使用WatchBP Office监测仪进行的测量进行了比较。采用配对t检验、Bland-Altman图和Pearson相关性分析来比较这两种设备。共分析了62例患者:他们的平均年龄为39.5±9.3岁,平均体重指数为45.1±5.7kg/m²。55例患者(89%)为女性,21例(34%)被诊断为高血压。BpTRU测量的平均血压值为117.2±12.7/76.7±9.7mmHg,WatchBP测量的平均血压值为126.6±13.8/79.7±10.9mmHg[收缩压差值为-9.4±11.6(P<0.0001),舒张压差值为-3.0±7.7mmHg(P=0.003)]。收缩压的Bland-Altman图显示出相当大的变异性。根据BpTRU,只有4例(6.5%)患者患有高血压(>140/90mmHg),而根据WatchBP则有15例(24%)。总之,使用两种常用的自动办公室设备测量的平均血压值存在显著差异,且差异具有临床重要意义。这些数据表明,需要对这些设备进行正式验证,尤其是在严重肥胖个体中。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验