Pfister C J, Harrison M A, Hamilton J W, Tompkins W J, Webster J G
IEEE Trans Biomed Eng. 1989 Apr;36(4):487-90. doi: 10.1109/10.18756.
We have developed a three-channel ambulatory esophageal pressure monitor and tested it with a series of 24-h studies. The monitor is a battery-operated, microprocessor-based device that measures pressures from three transducers positioned in the esophagus, stores the data in its memory, and transfers the data to an IBM PC computer system at the end of the recording period. Programs on the PC then analyze the data and identify contractile events, categorizing them according to specific parameters. Other programs display the pressure waveforms on the PC and allow visual inspection of the entire recordings or, alternatively, of particular events of interest. The system detects contractile abnormalities in patients with intermittent, noncardiac chest pains. We tested the system on ten normal subjects and found a relatively high incidence of what are usually considered "abnormal" contractions.
我们研发了一种三通道动态食管压力监测仪,并通过一系列24小时研究对其进行了测试。该监测仪是一种由电池供电、基于微处理器的设备,它可测量位于食管中的三个传感器的压力,将数据存储在其内存中,并在记录期结束时将数据传输到IBM个人计算机系统。个人计算机上的程序随后对数据进行分析,识别收缩事件,并根据特定参数对其进行分类。其他程序在个人计算机上显示压力波形,并允许对整个记录或感兴趣的特定事件进行目视检查。该系统可检测间歇性非心源性胸痛患者的收缩异常。我们在十名正常受试者身上测试了该系统,发现通常被认为“异常”的收缩发生率相对较高。