Gignoux C, Bonnet-Eymard P O, Hostein J, Fournet J
Gastroenterol Clin Biol. 1987 Jan;11(1):17-23.
Twenty-four hour pH monitoring is used increasingly as a technique for diagnosis of gastro-esophageal reflux (GER). The Synectics recording apparatus (Stockholm, Sweden) was used in 27 ambulatory control subjects in order to: identify the variations of the data-dependent factors (electrodes, buffer solutions, calibration, placement of esophageal electrode). The consequences of these variations were evaluated by studying the measurement deflection after 24 h, baseline variations, and a comparative double pH esophageal recording with reference equipment for 3 h postprandial; to provide normal values for several parameters of GER measurement under the conditions of this study (total number of reflux, number of reflux episodes longer than 5 min, duration of the longest reflux, percentage of time during which pH was less than 4.0). Important interindividual variations were observed. When interpreting the results provided by this type of apparatus, one should take into account these large variations due, first, to the technique used and second, to the conditions of ambulatory recording.
24小时pH监测作为诊断胃食管反流(GER)的一项技术,其应用越来越广泛。对27名动态对照受试者使用了Synectics记录设备(瑞典斯德哥尔摩),目的如下:确定数据相关因素(电极、缓冲溶液、校准、食管电极位置)的变化情况。通过研究24小时后的测量偏差、基线变化以及餐后3小时与参考设备进行的双pH食管记录对比,评估这些变化的影响;在本研究条件下提供GER测量若干参数的正常值(反流总数、持续时间超过5分钟的反流发作次数、最长反流持续时间、pH值小于4.0的时间百分比)。观察到了个体间的重要差异。在解释这类设备提供的结果时,首先应考虑到由于所使用的技术以及其次由于动态记录条件导致的这些巨大差异。