Hancu Ileana, Lee Seung-Kyun, Hulsey Keith, Lenkinski Robert, Holland Dominic, Sperl Jonathan I, Tan Ek T
GE Global Research Center, Niskayuna, NY, USA.
CNIR, IBS and Department of Biomedical Engineering, Sungkyunkwan University, Suwon, South Korea.
Magn Reson Med. 2017 Jul;78(1):247-253. doi: 10.1002/mrm.26328. Epub 2016 Jul 12.
To compare the effectiveness of prospective, retrospective, and combined (prospective + retrospective) EPI distortion correction methods in bilateral breast diffusion-weighted imaging (DWI) scans.
Five healthy female subjects underwent an axial bilateral breast DWI exam with and without prospective B inhomogeneity correction using slice-by-slice linear shimming. In each case, an additional b=0 DWI scan was performed with the polarity of the phase-encoding gradient reversed, to generate an estimated B map; this map or a separately acquired B map was used for retrospective correction, either alone or in combination with the prospective correction. The alignment between an undistorted, anatomical reference scan with similar contrast and the corrected b=0 DWI images with different correction schemes was assessed.
The average cross-correlation coefficient between the DWI images and the anatomical reference scan was increased from 0.82 to 0.92 over the five volunteers when combined prospective and retrospective distortion correction was applied. Furthermore, such correction substantially reduced patient-to-patient variation of the image alignment and the variability of the average apparent diffusion coefficient in normal glandular tissue.
Combined prospective and retrospective distortion correction can provide an efficient way to reduce susceptibility-induced image distortions and enhance the reliability of breast DWI exams. Magn Reson Med 78:247-253, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
比较前瞻性、回顾性以及联合(前瞻性 + 回顾性)回波平面成像(EPI)失真校正方法在双侧乳腺扩散加权成像(DWI)扫描中的有效性。
五名健康女性受试者接受了轴向双侧乳腺DWI检查,分别进行了使用逐片线性匀场的有无前瞻性B不均匀性校正的扫描。在每种情况下,通过反转相位编码梯度的极性进行额外的b = 0 DWI扫描,以生成估计的B图;该图或单独获取的B图用于单独或与前瞻性校正联合进行回顾性校正。评估具有相似对比度的未失真解剖参考扫描与采用不同校正方案的校正后b = 0 DWI图像之间的对齐情况。
当应用联合前瞻性和回顾性失真校正时,五名志愿者的DWI图像与解剖参考扫描之间的平均互相关系数从0.82提高到了0.92。此外,这种校正显著降低了图像对齐的患者间差异以及正常腺组织中平均表观扩散系数的变异性。
联合前瞻性和回顾性失真校正可以提供一种有效的方法来减少敏感性诱导的图像失真,并提高乳腺DWI检查的可靠性。《磁共振医学》78:247 - 253, 2017。© 2016国际磁共振医学学会。