Left arm/left leg lead reversals at the cable junction box: A cause for an epidemic of errors.

作者信息

Velagapudi Poonam, Turagam Mohit K, Ritter Sherry, Dohrmann Mary L

机构信息

Division of Cardiovascular Medicine, University of Missouri School of Medicine, Columbia, MO, USA.

Division of Cardiovascular Medicine, University of Missouri School of Medicine, Columbia, MO, USA.

出版信息

J Electrocardiol. 2017 Jan-Feb;50(1):111-114. doi: 10.1016/j.jelectrocard.2016.06.011. Epub 2016 Jul 5.

Abstract

Medical errors, especially due to misinterpretation of electrocardiograms (ECG), are extremely common in patients admitted to the hospital and significantly account for increased morbidity, mortality and health care costs in the United States. Inaccurate performance of an ECG can lead to invalid interpretation and in turn may lead to costly cardiovascular evaluation. We report a retrospective series of 58 sequential cases of ECG limb lead reversals in the ER due to inadvertent interchange in the lead cables at the point where they insert into the cable junction box of one ECG machine. This case series highlights recognition of ECG lead reversal originating in the ECG machine itself. This case series also demonstrates an ongoing need for education regarding standardization of ECG testing and for recognizing technical anomalies to deliver appropriate care for the patient.