Validation of a new closed circuit indirect calorimetry method compared with the open Douglas bag method.

作者信息

Raurich J M, Ibañez J, Marse P

机构信息

Intensive Care Unit, Hospital Son Dureta, Palma de Mallorca, Spain.

出版信息

Intensive Care Med. 1989;15(4):274-8. doi: 10.1007/BF00271067.

Abstract

New equipment designed for the routine measurement of oxygen uptake (VO2) using a closed circuit method has been validated by comparing it with a standard Douglas bag method. The equipment (The Caloric Measurement Unit, CMU) has been tested in 10 critically ill patients during mechanical ventilation (MV) and in 10 spontaneously breathing healthy subjects. Determinations of VO2 and of the resting energy expenditure (REE) were measured in duplicate with the standard method and once with the CMU. Six additional patients receiving MV were studied with the CMU to evaluate the reproducibility and the effect of FIO2 = 1 vs FIO2 = 0.43 on VO2 measurements. Considering the whole group of 10 patients and 10 subjects, the mean difference of VO2 between both methods was -2 +/- 21 ml/min (95% confidence interval, -11.8 to 7.8 ml/min, p = 0.6) standard deviation. Both methods had a similar reproducibility and the mean difference of VO2 measured at the two different FIO2 with the CMU was -3.2 +/- 11 ml/min (95% confidence interval, -14.7 to 8.4 ml/min, p = 0.5). No statistically significant difference was found between derived REE values obtained from either method. These data show a good correlation between the two methods suggesting that CMU may be used in place of the standard method with the same accuracy in measurement of VO2 even at FIO2 = 1.

摘要

一种设计用于采用闭路法常规测量氧摄取量(VO2)的新设备,已通过与标准道格拉斯袋法进行比较得到验证。该设备(热量测量单元,CMU)已在10例机械通气(MV)的重症患者和10例自主呼吸的健康受试者中进行了测试。采用标准方法对VO2和静息能量消耗(REE)进行了双份测定,并使用CMU进行了一次测定。另外6例接受MV的患者使用CMU进行研究,以评估重复性以及吸入氧分数(FIO2)=1与FIO2 = 0.43对VO2测量的影响。考虑到10例患者和10例受试者的整个组,两种方法之间VO2的平均差异为-2±21 ml/分钟(95%置信区间,-11.8至7.8 ml/分钟,p = 0.6)标准差。两种方法具有相似的重复性,并且使用CMU在两种不同FIO2下测量的VO2平均差异为-3.2±11 ml/分钟(95%置信区间,-14.7至8.4 ml/分钟,p = 0.5)。从两种方法获得的推导REE值之间未发现统计学上的显著差异。这些数据表明两种方法之间具有良好的相关性,这表明即使在FIO2 = 1时,CMU也可用于替代标准方法以相同的准确度测量VO2。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验