Porcelain surface conditioning protocols and shear bond strength of orthodontic brackets.

作者信息

Lestrade Ashley M, Ballard Richard W, Xu Xiaoming, Yu Qingzhao, Kee Edwin L, Armbruster Paul C

出版信息

Aust Orthod J. 2016 May;32(1):18-22.

Abstract

AIM

The objective of the present study was to determine which of six bonding protocols yielded a clinically acceptable shear bond strength (SBS) of metal orthodontic brackets to CAD/CAM lithium disilicate porcelain restorations. A secondary aim was to determine which bonding protocol produced the least surface damage at debond.

METHODS

Sixty lithium disilicate samples were fabricated to replicate the facial surface of a mandibular first molar using a CEREC CAD/CAM machine. The samples were split into six test groups, each of which received different mechanical/chemical pretreatment protocols to roughen the porcelain surface prior to bonding a molar orthodontic attachment. Shear bond strength testing was conducted using an Instron machine. The mean, maximum, minimal, and standard deviation SBS values for each sample group including an enamel control were calculated. A t-test was used to evaluate the statistical significance between the groups.

RESULTS

No significant differences were found in SBS values, with the exception of surface roughening with a green stone prior to HFA and silane treatment. This protocol yielded slightly higher bond strength which was statistically significant.

CONCLUSION

Chemical treatment alone with HFA/silane yielded SBS values within an acceptable clinical range to withstand forces applied by orthodontic treatment and potentially eliminates the need to mechanically roughen the ceramic surface.

摘要

目的

本研究的目的是确定六种粘结方案中哪一种能使金属正畸托槽与CAD/CAM二硅酸锂瓷修复体产生临床上可接受的剪切粘结强度(SBS)。第二个目的是确定哪种粘结方案在脱粘时产生的表面损伤最小。

方法

使用CEREC CAD/CAM机器制作60个二硅酸锂样本,以复制下颌第一磨牙的颊面。样本被分成六个测试组,每组在粘结磨牙正畸附件之前接受不同的机械/化学预处理方案,以使瓷表面粗糙。使用Instron机器进行剪切粘结强度测试。计算每个样本组(包括釉质对照组)的平均、最大、最小和标准偏差SBS值。使用t检验评估组间的统计学显著性。

结果

除了在氢氟酸(HFA)和硅烷处理之前用绿石进行表面粗化外,SBS值没有发现显著差异。该方案产生的粘结强度略高,具有统计学显著性。

结论

单独使用HFA/硅烷进行化学处理产生的SBS值在临床上可接受的范围内,足以承受正畸治疗施加的力,并且可能无需对陶瓷表面进行机械粗化。