Shin J H, Oestmann J, Hall D, Cardenosa G, McCarthy K A, Mrose H E, Pile-Spellman E, Rubens J R, Greene R E
Department of Radiology, Massachusetts General Hospital, Boston.
Radiology. 1989 Aug;172(2):399-401. doi: 10.1148/radiology.172.2.2748821.
The authors compared low-dose (32% of standard exposure) storage phosphor digital imaging (system resolution: 0.2-mm pixels, 10 bits) with isovoltage 75-kVp conventional radiography (standard exposure) in the detection of subtle simulated gastric abnormalities by using air contrast barium studies. Subtle simulated abnormalities (3-7-mm polyps, 4-15-mm ulcer craters, 4-11-mm-diameter edema, and 11-12-mm linear ulcers) were produced in resected canine stomachs. Receiver operating characteristic analysis of 1,800 observations by six readers indicated that the digital images with and without high-frequency edge enhancement were equivalent to conventional radiographs (mean receiver operating characteristic areas [+/- standard deviation]: 0.76 +/- 0.06, 0.78 +/- 0.04, and 0.77 +/- 0.04, respectively). The accuracy of the diagnosis was equivalent for all three modalities. The following mean accuracies of negative and positive responses, respectively, for unenhanced digital, edge-enhanced digital, and conventional images were determined: 0.71 +/- 0.05 and 0.41 +/- 0.07, 0.71 +/- 0.04 and 0.51 +/- 0.09, and 0.68 +/- 0.04 and 0.43 +/- 0.05. It was concluded that low-dose storage phosphor air-contrast barium studies were equivalent to conventional radiography in the detection of subtle gastric abnormalities.
作者将低剂量(标准曝光量的32%)存储磷光体数字成像(系统分辨率:0.2毫米像素,10位)与75 kVp等电压传统X线摄影(标准曝光量)用于气钡双重造影检查以检测细微的模拟胃部异常情况。在切除的犬胃中制造了细微的模拟异常(3 - 7毫米息肉、4 - 15毫米溃疡 crater、4 - 11毫米直径的水肿以及11 - 12毫米线性溃疡)。六位阅片者对1800次观察结果进行的受试者操作特征分析表明,有无高频边缘增强的数字图像与传统X线片相当(平均受试者操作特征面积[±标准差]:分别为0.76 ± 0.06、0.78 ± 0.04和0.77 ± 0.04)。三种检查方式的诊断准确性相当。确定了未增强数字图像、边缘增强数字图像和传统图像的阴性和阳性反应的平均准确率如下:0.71 ± 0.05和0.41 ± 0.07、0.71 ± 0.04和0.51 ± 0.09、0.68 ± 0.04和0.43 ± 0.05。得出的结论是,低剂量存储磷光体气钡双重造影检查在检测细微胃部异常方面与传统X线摄影相当。