Ferreira Tiago B, Ribeiro Filomena J
Department of Ophthalmology, Hospital da Luz, Lisbon, Portugal.
Clin Ophthalmol. 2016 Aug 13;10:1521-9. doi: 10.2147/OPTH.S113027. eCollection 2016.
To assess the accuracy of corneal astigmatism evaluation measured by four techniques, Orbscan IIz(®), Lenstar LS900(®), Cassini(®), and Total Cassini (anterior + posterior surface), in pseudophakic eyes.
A total of 30 patients (46 eyes) who had undergone cataract surgery with the implantation of a monofocal intraocular lens (AcrySof IQ) were assessed after surgery. For each eye, subjective assessment of astigmatism and its axis was performed. Minimum, maximum, and mean keratometry and astigmatism and its axis were evaluated using the four measurement techniques. All measurements were compared with the subjective measurements. Agreement between each measurement technique and subjective assessment was evaluated using Bland-Altman plots. Linear regressions were performed and compared.
Linear regression analysis of astigmatism axis showed very high R (2) for all models, with Total Cassini showing the least difference to the unit slope (0.052) and the least difference to a null constant (3.790), although not statistically different from the other models. Regarding astigmatism value, the Cassini and Total Cassini models were similar and statistically better than the Lenstar model. Cassini and Total Cassini showed better J0 compared with Orbscan.
On linear regression models, Cassini and Total Cassini showed the best performance regarding astigmatism value. Cassini and Total Cassini also showed the least J0 deviation from the Cartesian origin compared with Orbscan, which had the lowest performance. Total corneal measurement with the color LED topographer seems to be a better technique for astigmatism assessment.
评估Orbscan IIz(®)、Lenstar LS900(®)、Cassini(®)和Total Cassini(前后表面)这四种技术测量人工晶状体眼角膜散光的准确性。
对30例(46只眼)接受白内障手术并植入单焦点人工晶状体(AcrySof IQ)的患者术后进行评估。对每只眼进行散光及其轴位的主观评估。使用这四种测量技术评估最小、最大和平均角膜曲率、散光及其轴位。所有测量结果均与主观测量结果进行比较。使用Bland-Altman图评估每种测量技术与主观评估之间的一致性。进行线性回归并比较。
散光轴位的线性回归分析显示,所有模型的R(2)值都非常高,Total Cassini与单位斜率的差异最小(0.052),与零常数的差异最小(3.790),尽管与其他模型无统计学差异。关于散光值,Cassini和Total Cassini模型相似,且在统计学上优于Lenstar模型。与Orbscan相比,Cassini和Total Cassini的J0值更好。
在线性回归模型中,Cassini和Total Cassini在散光值方面表现最佳。与性能最低的Orbscan相比,Cassini和Total Cassini与笛卡尔原点的J0偏差也最小。使用彩色LED地形图仪进行全角膜测量似乎是评估散光的更好技术。