两种不同放射治疗图像引导系统的机械和几何精度评估。

Evaluation of mechanical and geometric accuracy of two different image guidance systems in radiotherapy.

作者信息

Kanakavelu Nithya, Ravindran Anand Mambakam, Samuel Emmanvelrajan James Jebaseelan

机构信息

Department of Radiation Oncology, Medwin Cancer Center, Hyderabad, India; Photonics, Nuclear and Medical Physics Division, VIT University, Vellore, India.

Department of Radiation Oncology, SevenHills Hospital, Mumbai, India.

出版信息

Rep Pract Oncol Radiother. 2016 May-Jun;21(3):259-65. doi: 10.1016/j.rpor.2015.11.005. Epub 2016 Jan 4.

Abstract

AIM

To assess the mechanical and the geometric accuracy of two different clinically used image guidance systems in radiotherapy for a period of 6 months.

BACKGROUND

With the image guidance procedures being routine in the clinical radiotherapy department, the quality assurance tests for these systems become essential. The mechanical and geometric accuracy of these systems are crucial since it directly affects patient treatment set-up and delivery.

MATERIALS AND METHODS

We have assessed the mechanical and the geometric accuracy of two different image guidance systems (MV and kV based), being used clinically for a period of 6 months. The quality assurance tests such as imager positioning/repositioning, imaging and treatment beam isocentre coincidence, imager mechanical alignment, image scaling, geometric accuracy of cone beam computed tomography system, automatic image registration and offset calculation accuracy were assessed in this period.

RESULTS

It was found that both systems were mechanically and geometrically accurate within ±2 mm in this period.

CONCLUSION

The quality assurance tests for MV based image guidance system were simple compared to kV based systems. We recommend performing periodic quality assurance tests to verify the integrity of both image guidance systems.

摘要

目的

评估两种不同临床使用的放射治疗图像引导系统在6个月期间的机械和几何精度。

背景

随着图像引导程序在临床放射治疗科室成为常规操作,对这些系统的质量保证测试变得至关重要。这些系统的机械和几何精度至关重要,因为它直接影响患者治疗的设置和实施。

材料与方法

我们评估了两种不同的图像引导系统(基于MV和kV)在临床上使用6个月期间的机械和几何精度。在此期间评估了诸如成像仪定位/重新定位、成像和治疗束等中心重合、成像仪机械对准、图像缩放、锥形束计算机断层扫描系统的几何精度、自动图像配准和偏移计算精度等质量保证测试。

结果

在此期间发现两种系统在机械和几何方面的精度均在±2毫米以内。

结论

与基于kV的系统相比,基于MV的图像引导系统的质量保证测试更简单。我们建议定期进行质量保证测试以验证两种图像引导系统的完整性。

相似文献

3
Assessment and evaluation of MV image guidance system performance in radiotherapy.放射治疗中MV图像引导系统性能的评估与评价。
Rep Pract Oncol Radiother. 2015 May-Jun;20(3):188-97. doi: 10.1016/j.rpor.2015.01.002. Epub 2015 Feb 7.

引用本文的文献

2
Registration methods in radiotherapy.放射治疗中的定位方法。
Rep Pract Oncol Radiother. 2019 Jan-Feb;24(1):28-34. doi: 10.1016/j.rpor.2018.09.004. Epub 2018 Oct 10.

本文引用的文献

2
Assessment and evaluation of MV image guidance system performance in radiotherapy.放射治疗中MV图像引导系统性能的评估与评价。
Rep Pract Oncol Radiother. 2015 May-Jun;20(3):188-97. doi: 10.1016/j.rpor.2015.01.002. Epub 2015 Feb 7.
4
Image guidance in radiation therapy: techniques and applications.放射治疗中的图像引导:技术与应用
Radiol Res Pract. 2014;2014:705604. doi: 10.1155/2014/705604. Epub 2014 Dec 17.
7
Comparative analysis of image guidance in two institutions for prostate cancer patients.两家机构对前列腺癌患者进行图像引导的对比分析。
Rep Pract Oncol Radiother. 2014 Jan 2;19(3):206-13. doi: 10.1016/j.rpor.2013.12.001. eCollection 2014 May.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验