Kanakavelu Nithya, Ravindran Anand Mambakam, Samuel Emmanvelrajan James Jebaseelan
Department of Radiation Oncology, Medwin Cancer Center, Hyderabad, India; Photonics, Nuclear and Medical Physics Division, VIT University, Vellore, India.
Department of Radiation Oncology, SevenHills Hospital, Mumbai, India.
Rep Pract Oncol Radiother. 2016 May-Jun;21(3):259-65. doi: 10.1016/j.rpor.2015.11.005. Epub 2016 Jan 4.
To assess the mechanical and the geometric accuracy of two different clinically used image guidance systems in radiotherapy for a period of 6 months.
With the image guidance procedures being routine in the clinical radiotherapy department, the quality assurance tests for these systems become essential. The mechanical and geometric accuracy of these systems are crucial since it directly affects patient treatment set-up and delivery.
We have assessed the mechanical and the geometric accuracy of two different image guidance systems (MV and kV based), being used clinically for a period of 6 months. The quality assurance tests such as imager positioning/repositioning, imaging and treatment beam isocentre coincidence, imager mechanical alignment, image scaling, geometric accuracy of cone beam computed tomography system, automatic image registration and offset calculation accuracy were assessed in this period.
It was found that both systems were mechanically and geometrically accurate within ±2 mm in this period.
The quality assurance tests for MV based image guidance system were simple compared to kV based systems. We recommend performing periodic quality assurance tests to verify the integrity of both image guidance systems.
评估两种不同临床使用的放射治疗图像引导系统在6个月期间的机械和几何精度。
随着图像引导程序在临床放射治疗科室成为常规操作,对这些系统的质量保证测试变得至关重要。这些系统的机械和几何精度至关重要,因为它直接影响患者治疗的设置和实施。
我们评估了两种不同的图像引导系统(基于MV和kV)在临床上使用6个月期间的机械和几何精度。在此期间评估了诸如成像仪定位/重新定位、成像和治疗束等中心重合、成像仪机械对准、图像缩放、锥形束计算机断层扫描系统的几何精度、自动图像配准和偏移计算精度等质量保证测试。
在此期间发现两种系统在机械和几何方面的精度均在±2毫米以内。
与基于kV的系统相比,基于MV的图像引导系统的质量保证测试更简单。我们建议定期进行质量保证测试以验证两种图像引导系统的完整性。