不同研究方案下应用碲锌镉和闪烁相机行心肌灌注 SPECT 的比较。

Comparison between stress myocardial perfusion SPECT recorded with cadmium-zinc-telluride and Anger cameras in various study protocols.

机构信息

Department of Nuclear Medicine, CHU-Nancy, Nancy, 54000, France.

出版信息

Eur J Nucl Med Mol Imaging. 2013 Feb;40(3):331-40. doi: 10.1007/s00259-012-2292-8. Epub 2012 Nov 27.

Abstract

PURPOSE

The results of stress myocardial perfusion SPECT could be enhanced by new cadmium-zinc-telluride (CZT) cameras, although differences compared to the results with conventional Anger cameras remain poorly known for most study protocols. This study was aimed at comparing the results of CZT and Anger SPECT according to various study protocols while taking into account the influence of obesity.

METHODS

The study population, which was from three different institutions equipped with identical CZT cameras, comprised 276 patients referred for study using protocols involving (201)Tl (n = 120) or (99m)Tc-sestamibi injected at low dose at stress ((99m)Tc-Low; stress/rest 1-day protocol; n = 110) or at high dose at stress ((99m)Tc-High; rest/stress 1-day or 2-day protocol; n = 46). Each Anger SPECT scan was followed by a high-speed CZT SPECT scan (2 to 4 min).

RESULTS

Agreement rates between CZT and Anger SPECT were good irrespective of the study protocol (for abnormal SPECT, (201)Tl 92 %, (99m)Tc-Low 86 %, (99m)Tc-High 98 %), although quality scores were much higher for CZT SPECT with all study protocols. Overall correlations were high for the extent of myocardial infarction (r = 0.80) and a little lower for ischaemic areas (r = 0.72), the latter being larger on Anger SPECT (p < 0.001). This larger extent was mainly observed in 50 obese patients who were in the (201)Tl or (99m)Tc-Low group and in whom stress myocardial counts were particularly low with Anger SPECT (228 ± 101 kcounts) and dramatically enhanced with CZT SPECT (+279 ± 251 %).

CONCLUSION

Concordance between the results of CZT and Anger SPECT is good regardless of study protocol and especially when excluding obese patients who have low-count Anger SPECT and for whom myocardial counts are dramatically enhanced on CZT SPECT.

摘要

目的

新型碲锌镉(CZT)相机可增强应激心肌灌注 SPECT 的结果,尽管对于大多数研究方案而言,与传统的 Anger 相机相比,其结果差异仍知之甚少。本研究旨在比较 CZT 和 Anger SPECT 的结果,同时考虑肥胖的影响,根据不同的研究方案进行比较。

方法

研究人群来自三个配备相同 CZT 相机的机构,包括 276 名因方案而接受检查的患者,这些方案涉及(201)Tl(n=120)或低剂量(99m)Tc-甲氧基异丁基异腈(99m)Tc-Low(应激/休息 1 天方案;n=110)或高剂量(99m)Tc-High(1 天或 2 天应激/休息方案;n=46)。每个 Anger SPECT 扫描后立即进行高速 CZT SPECT 扫描(2 至 4 分钟)。

结果

无论研究方案如何(对于异常 SPECT,(201)Tl 92%,(99m)Tc-Low 86%,(99m)Tc-High 98%),CZT 和 Anger SPECT 的一致性率均良好,尽管所有研究方案的 CZT SPECT 的质量评分都高得多。对于心肌梗死的范围(r=0.80)和缺血区的相关性较高(r=0.72),后者在 Anger SPECT 上更大(p<0.001)。这种较大的范围主要发生在 50 名肥胖患者中,这些患者在(201)Tl 或(99m)Tc-Low 组中,并且在用 Anger SPECT 时心肌计数特别低(228±101 kcounts),而用 CZT SPECT 则明显增强(+279±251 kcounts)。

结论

无论研究方案如何,CZT 和 Anger SPECT 的结果一致性都很好,尤其是在排除了那些低计数的 Anger SPECT 且 CZT SPECT 心肌计数明显增强的肥胖患者时。