J Nucl Cardiol. 2021 Aug;28(4):1323-1330. doi: 10.1007/s12350-019-01835-4. Epub 2019 Aug 5.
To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI).
We included 19 patients that underwent sequential low-dose Tc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer's uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared.
The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen's κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98).
Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.
比较配备 CZT 探测器的新型通用 SPECT 相机(Discovery NM/CT 670)与专用 CZT(Alcyone、Discovery 530c)心脏相机在接受心肌灌注成像(MPI)的患者中的表现。
我们纳入了 19 名连续接受低剂量 Tc-四氮茂(负荷时 148-185MBq,静息时 296-370MBq)MPI 的患者,分别使用 Alcyone 和 Discovery 670 相机进行检查。对每个扫描进行定量(%示踪剂摄取)和半定量灌注数据分析。此外,从每个相机获取主要左心室(LV)功能和结构参数并进行比较。
两台相机在负荷时(R=0.90,P<0.001)和静息时(R=0.88,P<0.001)的节段性心肌%摄取方面显示出极好的相关性,且 Bland-Altman 一致性界限较窄。两台相机在灌注分析方面的诊断一致性水平非常高(Cohen's κ=0.85)。同样,两台相机在 LV 射血分数(R=0.95)、峰值充盈率(R=0.97)和质量(R=0.98)的评估方面也显示出极好的相关性。
我们的初步结果表明,使用通用的 Discovery 670 CZT-SPECT 相机进行 MPI 是可行的,与当前的最先进技术相比表现良好。