使用光野的万向直线加速器系统进行动态肿瘤追踪的简单质量保证方法。

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field.

作者信息

Miura Hideharu, Ozawa Shuichi, Hayata Masahiro, Tsuda Shintaro, Yamada Kiyoshi, Nagata Yasushi

机构信息

Hiroshima High-Precision Radiotherapy Cancer Center.

出版信息

J Appl Clin Med Phys. 2016 Sep 8;17(5):177-183. doi: 10.1120/jacmp.v17i5.6376.

Abstract

We proposed a simple visual method for evaluating the dynamic tumor tracking (DTT) accuracy of a gimbal mechanism using a light field. A single photon beam was set with a field size of 30 × 30 mm2 at a gantry angle of 90°. The center of a cube phantom was set up at the isocenter of a motion table, and 4D modeling was performed based on the tumor and infrared (IR) marker motion. After 4D modeling, the cube phantom was replaced with a sheet of paper, which was placed perpen-dicularly, and a light field was projected on the sheet of paper. The light field was recorded using a web camera in a treatment room that was as dark as possible. Calculated images from each image obtained using the camera were summed to compose a total summation image. Sinusoidal motion sequences were produced by moving the phantom with a fixed amplitude of 20 mm and different breathing periods of 2, 4, 6, and 8 s. The light field was projected on the sheet of paper under three conditions: with the moving phantom and DTT based on the motion of the phantom, with the moving phantom and non-DTT, and with a stationary phantom for comparison. The values of tracking errors using the light field were 1.12 ± 0.72, 0.31 ± 0.19, 0.27 ± 0.12, and 0.15 ± 0.09 mm for breathing periods of 2, 4, 6, and 8s, respectively. The tracking accuracy showed dependence on the breath-ing period. We proposed a simple quality assurance (QA) process for the tracking accuracy of a gimbal mechanism system using a light field and web camera. Our method can assess the tracking accuracy using a light field without irradiation and clearly visualize distributions like film dosimetry.

摘要

我们提出了一种简单的视觉方法,用于评估使用光场的万向节机构的动态肿瘤跟踪(DTT)精度。在90°的机架角度下,设置了一个场大小为30×30 mm2的单光子束。将立方体模体的中心设置在运动台的等中心处,并基于肿瘤和红外(IR)标记物的运动进行4D建模。4D建模后,将立方体模体换成一张垂直放置的纸,并将光场投射到这张纸上。使用治疗室内尽可能暗的网络摄像头记录光场。对使用摄像头获得的每个图像的计算图像进行求和,以组成一个总求和图像。通过以20 mm的固定振幅和2、4、6和8 s的不同呼吸周期移动模体来生成正弦运动序列。在三种条件下将光场投射到纸上:有移动模体且基于模体运动的DTT、有移动模体且无DTT、以及有固定模体用于比较。对于2、4、6和8 s的呼吸周期,使用光场的跟踪误差值分别为1.12±0.72、0.31±0.19、0.27±0.12和0.15±0.09 mm。跟踪精度显示出对呼吸周期的依赖性。我们提出了一种使用光场和网络摄像头对万向节机构系统的跟踪精度进行简单质量保证(QA)的过程。我们的方法可以在不进行照射的情况下使用光场评估跟踪精度,并能像胶片剂量测定法一样清晰地可视化分布情况。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验