Eghtedari Mohammad, Ma Jingfei, Fox Patricia, Guvenc Inanc, Yang Wei T, Dogan Basak E
Department of Radiology, University of California San Diego, San Diego, CA, USA.
Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA.
Quant Imaging Med Surg. 2016 Aug;6(4):374-380. doi: 10.21037/qims.2016.07.06.
To evaluate the effect of b value or the magnetic field strength (B0) on the sensitivity and specificity of quantitative breast diffusion-weighted imaging (DWI).
A total of 126 patients underwent clinical breast MRI that included pre-contrast DWI imaging using b values of both 1,000 and 1,500 s/mm at either 1.5 T (n=86) or 3.0 T (n=40). Quantitative apparent diffusion coefficients (ADC) were measured and compared for 18 benign, 33 malignant lesions, and 126 normal breast tissues. Optimal ADCmean threshold for differentiating benign and malignant lesions was estimated and the effect of b values and B0 were examined using a generalized estimating equations (GEE) model.
The optimal ADCmean threshold was 1.235×10 mm/s for b value of 1,000 and 0.934×10 mm/s for b value of 1,500. Using these thresholds, the sensitivities and specificities were 96% and 89% (b value =1,000, B0 =1.5 T), 89% and 98% (b value =1,000, B0 =3.0 T), 88% and 96% (b value =1,500, B0 =1.5 T), and 67% and 100% (b value =1,500, B0 =3.0 T). No significant difference was found between different B0 (P=0.26) or b values (P=0.28).
Better sensitivity is achieved with DWI of b value =1,000 than with b value =1,500. However, b value and B0 do not significantly impact diagnostic performance of DWI when using appropriate thresholds.
评估b值或磁场强度(B0)对乳腺定量扩散加权成像(DWI)敏感性和特异性的影响。
126例患者接受了乳腺临床MRI检查,其中包括在1.5 T(n = 86)或3.0 T(n = 40)场强下使用1000和1500 s/mm²的b值进行的对比前DWI成像。对18个良性病变、33个恶性病变及126个正常乳腺组织测量并比较了定量表观扩散系数(ADC)。估计了鉴别良恶性病变的最佳平均ADC阈值,并使用广义估计方程(GEE)模型研究了b值和B0的影响。
b值为1000时最佳平均ADC阈值为1.235×10⁻³mm²/s,b值为1500时为0.934×10⁻³mm²/s。使用这些阈值时,敏感性和特异性分别为:b值 = 1000,B0 = 1.5 T时为96%和89%;b值 = 1000,B0 = 3.0 T时为89%和98%;b值 = 1500,B0 = 1.5 T时为88%和96%;b值 = 1500,B0 = 3.0 T时为67%和100%。不同B0(P = 0.26)或b值(P = 0.28)之间未发现显著差异。
b值 = 1000的DWI比b值 = 1500的DWI具有更高的敏感性。然而,当使用适当阈值时,b值和B0对DWI的诊断性能无显著影响。