The conical stent in coronary artery improves hemodynamics compared with the traditional cylindrical stent.

作者信息

Yu Yi, Zhou Yujie, Ma Qian, Jia Shuo, Wu Sijing, Sun Yan, Liu Xiaoli, Zhao Yingxin, Liu Yuyang, Shi Dongmei

机构信息

Beijing Anzhen Hospital, Capital Medical University, Beijing Institute of Heart Lung and Blood Vessel Disease, The Key Laboratory of Remodeling-related Cardiovascular Disease, Ministry of Education, Beijing 100029, China.

Beijing Anzhen Hospital, Capital Medical University, Beijing Institute of Heart Lung and Blood Vessel Disease, The Key Laboratory of Remodeling-related Cardiovascular Disease, Ministry of Education, Beijing 100029, China.

出版信息

Int J Cardiol. 2017 Jan 15;227:166-171. doi: 10.1016/j.ijcard.2016.11.065. Epub 2016 Nov 8.

Abstract

OBJECTIVES

This study sought to explore the efficacy of the conical stent implantation in the coronary artery by comparing the effects of cylindrical and conical stents on wall shear stress (WSS) and velocity of flow and fractional flow reserve (FFR).

BACKGROUND

The traditional cylindrical stent currently used in the percutaneous coronary intervention (PCI) has a consistent diameter, which does not match the physiological change of the coronary artery. On the contrary, as a new patent, the conical stent with tapering lumen is consistent with the physiological change of vascular diameter. However, the effect of the conical stent implantation on the coronary hemodynamics remains unclear.

METHODS

The coronary artery, artery stenosis and two stent models were established by Solidworks software. All models were imported into the computational fluid dynamics (CFD) software ANSYS ICEM-CFD to establish the fluid model. After the boundary conditions were set, CFD analysis was proceeded to compare the effects of two stent implantation on the change of WSS, velocity of flow and FFR.

RESULTS

Hemodynamic indexes including FFR, blood flow velocity distribution (BVD) and WSS were improved by either the cylindrical or the conical stent implantation. However, after the conical stent implantation, the change of FFR seemed to be slower and more homogenous; the blood flow velocity was more appropriate without any obvious blood stagnation and direction changes; the WSS after the conical stent implantation was uniform from the proximal to distal side of the stent.

CONCLUSIONS

Compared with the cylindrical stent, the conical stent implantation in the coronary artery can make the changes of vascular hemodynamic more closer to the physiological condition, which can reduce the incidence of intra-stent restenosis and thrombosis, thus making it more suitable for PCI therapy.

摘要

目的

本研究旨在通过比较圆柱形和圆锥形支架对壁面剪应力(WSS)、血流速度和血流储备分数(FFR)的影响,探讨冠状动脉内圆锥形支架植入的疗效。

背景

目前经皮冠状动脉介入治疗(PCI)中使用的传统圆柱形支架直径一致,与冠状动脉的生理变化不匹配。相反,作为一项新专利,具有逐渐变细内腔的圆锥形支架与血管直径的生理变化一致。然而,圆锥形支架植入对冠状动脉血流动力学的影响仍不清楚。

方法

利用Solidworks软件建立冠状动脉、动脉狭窄和两种支架模型。将所有模型导入计算流体动力学(CFD)软件ANSYS ICEM-CFD中建立流体模型。设置边界条件后,进行CFD分析,比较两种支架植入对WSS、血流速度和FFR变化的影响。

结果

圆柱形和圆锥形支架植入均可改善包括FFR、血流速度分布(BVD)和WSS在内的血流动力学指标。然而,圆锥形支架植入后,FFR的变化似乎更缓慢且更均匀;血流速度更合适,没有明显的血流停滞和方向改变;圆锥形支架植入后的WSS从支架近端到远端是均匀的。

结论

与圆柱形支架相比,冠状动脉内植入圆锥形支架可使血管血流动力学变化更接近生理状态,从而降低支架内再狭窄和血栓形成的发生率,使其更适合PCI治疗。