Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.

作者信息

Miura Hideharu, Ozawa Shuichi, Hosono Fumika, Sumida Naoki, Okazue Toshiya, Yamada Kiyoshi, Nagata Yasushi

机构信息

Hiroshima High-Precision Radiotherapy Cancer Center.

出版信息

J Appl Clin Med Phys. 2016 Nov 8;17(6):312-322. doi: 10.1120/jacmp.v17i6.6281.

Abstract

Radiochromic films are important tools for assessing complex dose distributions. Gafchromic EBT-XD films have been designed for optimal performance in the 40-4,000 cGy dose range. We investigated the dosimetric characteristics of these films, including their dose-response, postexposure density growth, and dependence on scanner orientation, beam energy, and dose rate with applications to high-dose volumetric-modulated arc therapy (VMAT) verification. A 10 MV beam from a TrueBeam STx linear accelerator was used to irradiate the films with doses in the 0-4,000 cGy range. Postexposure coloration was analyzed at postirradiation times ranging from several minutes to 48 h. The films were also irradiated with 6 MV (dose rate (DR): 600 MU/min), 6 MV flattening filter-free (FFF) (DR: 1,400 MU/ min), and 10 MV FFF (DR: 2,400 MU/min) beams to determine the energy and dose-rate dependence. For clinical examinations, we compared the dose distribu-tion measured with EBT-XD films and calculated by the planning system for four VMAT cases. The red channel of the EBT-XD film exhibited a wider dynamic range than the green and blue channels. Scanner orientation yielded a variation of ~ 3% in the net optical density (OD). The difference between the film front and back scan orientations was negligible, with variation of ~ 1.3% in the net OD. The net OD increased sharply within the first 6 hrs after irradiation and gradually afterwards. No significant difference was observed for the beam energy and dose rate, with a variation of ~ 1.5% in the net OD. The gamma passing rates (at 3%, 3 mm) between the film- measured and treatment planning system (TPS)-calculated dose distributions under a high dose VMAT plan in the absolute dose mode were more than 98.9%.

摘要

放射变色胶片是评估复杂剂量分布的重要工具。Gafchromic EBT-XD胶片专为在40-4000 cGy剂量范围内实现最佳性能而设计。我们研究了这些胶片的剂量学特性,包括它们的剂量响应、照射后密度增长以及对扫描仪方向、束流能量和剂量率的依赖性,并将其应用于高剂量容积调强弧形治疗(VMAT)验证。使用TrueBeam STx直线加速器的10 MV束流对胶片进行0-4000 cGy范围内的剂量照射。在照射后几分钟至48小时的时间范围内分析照射后的颜色变化。还使用6 MV(剂量率(DR):600 MU/分钟)、6 MV无均整器(FFF)(DR:1400 MU/分钟)和10 MV FFF(DR:2400 MU/分钟)束流对胶片进行照射,以确定能量和剂量率依赖性。对于临床检查,我们比较了用EBT-XD胶片测量的和计划系统计算的四个VMAT病例的剂量分布。EBT-XD胶片的红色通道比绿色和蓝色通道表现出更宽的动态范围。扫描仪方向导致净光密度(OD)变化约3%。胶片前后扫描方向之间的差异可忽略不计,净OD变化约1.3%。净OD在照射后的前6小时内急剧增加,之后逐渐增加。未观察到束流能量和剂量率有显著差异,净OD变化约1.5%。在绝对剂量模式下,高剂量VMAT计划下胶片测量的和治疗计划系统(TPS)计算的剂量分布之间的γ通过率(3%,3毫米)超过98.9%。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/be1e/5690510/e01e5a8a81ea/ACM2-17-312-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验