Palmer Antony L, Dimitriadis Alexis, Nisbet Andrew, Clark Catharine H
Department of Physics, Faculty of Engineering and Physical Science, University of Surrey, GU2 7JP, UK. Medical Physics Department, F-Level, Queen Alexandra Hospital, Portsmouth Hospitals NHS Trust, Portsmouth, PO6 3LY, UK.
Phys Med Biol. 2015 Nov 21;60(22):8741-52. doi: 10.1088/0031-9155/60/22/8741. Epub 2015 Oct 29.
There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison.
与治疗计划系统的计算相比,对于复杂治疗(尤其是小射野)的剂量输送验证,胶片剂量学重新引起了人们的兴趣。一种新型放射变色胶片Gafchromic EBT-XD可用于高剂量治疗验证,我们首次发表了对其使用的评估。我们评估了这种新型胶片用于兆伏级光子剂量学的性能,包括校准曲线、单通道和三通道剂量学性能以及与现有EBT3胶片的比较。在验证典型的25 Gy立体定向放射治疗(SRS)时,与TPS计划的剂量分布相比,使用三通道剂量学的EBT-XD在等剂量叠加方面显示出极佳的一致性,在200 - 2400 cGy范围内最大差异为1.0毫米,在γ评估中,在3%局部归一化、1.5毫米标准下平均通过率为97%。与EBT3相比,EBT-XD对SRS计划给出了更好的评估结果,在高剂量下校准曲线梯度有所改善,并且横向扫描仪效应降低。两种胶片尺寸相同。相同剂量下EBT-XD的光学密度低于EBT3。在兆伏级放射治疗中,两者的有效原子序数均可视为水等效。我们已经验证了EBT-XD在高剂量、小射野放射治疗中的应用,可用于常规质量控制以及即将进行的多中心SRS剂量学比对。