Aliseda Alberto, Chivukula Venkat Keshav, Mcgah Patrick, Prisco Anthony R, Beckman Jennifer A, Garcia Guilherme J M, Mokadam Nahush A, Mahr Claudius
From the *Department of Mechanical Engineering, University of Washington, Seattle, Washington; †Biotechnology and Bioengineering Center, Medical College of Wisconsin, Milwaukee, Wisconsin; ‡Division of Cardiology, University of Washington, Seattle, Washington; and §Division of Cardiothoracic Surgery, University of Washington, Seattle, Washington.
ASAIO J. 2017 Jan/Feb;63(1):14-23. doi: 10.1097/MAT.0000000000000443.
This study quantifies thrombogenic potential (TP) of a wide range of left ventricular assist device (LVAD) outflow graft anastomosis angles through state-of-the-art techniques: 3D imaged-based patient-specific models created via virtual surgery and unsteady computational fluid dynamics with Lagrangian particle tracking. This study aims at clarifying the influence of a single parameter (outflow graft angle) on the thrombogenesis associated with flow patterns in the aortic root after LVAD implantation. This is an important and poorly-understood aspect of LVAD therapy, because several studies have shown strong inter and intrapatient thrombogenic variability and current LVAD implantation strategies do not incorporate outflow graft angle optimization. Accurate platelet-level investigation, enabled by statistical treatment of outliers in Lagrangian particle tracking, demonstrates a strong influence of outflow graft anastomoses angle on thrombogenicity (platelet residence times and activation state characterized by shear stress accumulation) with significantly reduced TP for acutely-angled anastomosed outflow grafts. The methodology presented in this study provides a device-neutral platform for conducting comprehensive thrombogenicity evaluation of LVAD surgical configurations, empowering optimal patient-focused surgical strategies for long-term treatment and care for advanced heart failure patients.
本研究通过先进技术对多种左心室辅助装置(LVAD)流出道移植物吻合角度的血栓形成潜力(TP)进行了量化:通过虚拟手术创建基于3D成像的患者特异性模型,并结合拉格朗日粒子跟踪的非定常计算流体动力学。本研究旨在阐明单个参数(流出道移植物角度)对LVAD植入后主动脉根部血流模式相关血栓形成的影响。这是LVAD治疗中一个重要且尚未得到充分理解的方面,因为多项研究表明患者之间和患者自身的血栓形成存在很大差异,并且目前的LVAD植入策略并未纳入流出道移植物角度优化。通过对拉格朗日粒子跟踪中的异常值进行统计处理实现的精确血小板水平研究表明,流出道移植物吻合角度对血栓形成性(以剪切应力积累为特征的血小板停留时间和激活状态)有很大影响,急性角度吻合的流出道移植物的TP显著降低。本研究中提出的方法为LVAD手术配置的全面血栓形成性评估提供了一个与设备无关的平台,为晚期心力衰竭患者的长期治疗和护理提供了以患者为中心的最佳手术策略。