Krizsan Aron K, Kukuts Kornel, Al-Muhanna Walid, Szoboszlai Zoltan, Balazs Laszlo, Szabo Balazs, Kiss Janos, Nekolla Stephan, Barna Sandor, Garai Ildiko, Bukki Tamas, Forgacs Attila
ScanoMed Nuclear Medicine Centers, Nagyerdei Krt. 98, Debrecen, 4032, Hungary.
Mediso Ltd., Budapest, Hungary.
EJNMMI Phys. 2023 Mar 25;10(1):24. doi: 10.1186/s40658-023-00541-y.
In this study we evaluated the imaging capabilities of a novel Multi-pinhole collimator (MPH-Cardiac) specially designed for nuclear cardiology imaging on a Triple-NaI-detector based SPECT/CT system.
Tc point source measurements covering the field of view (FOV) were used to determine tomographic sensitivity (TS) and spatial resolution. Organ-size tomographic sensitivity (TS) was measured with a left ventricle (LV) phantom filled with typical myocardial activity of a patient scan. Reconstructed image uniformity was measured with a 140 mm diameter uniform cylinder phantom. Using the LV phantom once filled with Tc and after with I, Contrast-to-noise ratio (CNR) was measured on the reconstructed images by ROI analysis on the myocardium activity and on the LV cavity. Furthermore, a polar map analysis was performed determining Spill-Over-Ratio in water (SOR) and image noise. The results were compared with that of a dual-head parallel-hole low energy high resolution (LEHR) collimator system. A patient with suspected coronary artery disease (CAD) was scanned on the LEHR system using local protocol of 16 min total acquisition time, followed by a 4-min MPH-Cardiac scan.
Peak TS was found to be 1013 cps/MBq in the axial center of the FOV while it was decreasing toward the radial edges. TS in the CFOV was found to be 134 cps/MBq and 700 cps/MBq for the LEHR and MPH-Cardiac, respectively. Average spatial resolution throughout the FOV was 4.38 mm FWHM for the MPH-Cardiac collimator. Reconstructed image uniformity values were found to be 0.292% versus 0.214% for the LEHR and MPH-Cardiac measurements, respectively. CNR was found to be higher in case of MPH-Cardiac than for LEHR in case of Tc (15.5 vs. 11.7) as well as for I (13.5 vs. 8.3). SOR values were found to be 28.83% and 21.1% for the Tc measurements, and 31.44% and 24.33% for the I measurements for LEHR and MPH-Cardiac, respectively. Pixel noise of the Tc polar maps resulted in values of 0.38% and 0.24% and of the I polar maps 0.62% and 0.21% for LEHR and MPH-Cardiac, respectively. Visually interpreting the patient scan images, MPH-Cardiac resulted in better image contrast compared to the LEHR technique with four times shorter scan duration.
The significant image quality improvement achieved with dedicated MPH-Cardiac collimator on triple head SPECT/CT system paves the way for short acquisition and low-dose cardiovascular SPECT applications.
在本研究中,我们评估了一种专门为核心脏病学成像设计的新型多针孔准直器(MPH - 心脏型)在基于三碘化钠探测器的SPECT/CT系统上的成像能力。
使用覆盖视野(FOV)的锝点源测量来确定断层灵敏度(TS)和空间分辨率。使用填充有患者扫描典型心肌活性的左心室(LV)模型测量器官大小断层灵敏度(TS)。使用直径140毫米的均匀圆柱体模型测量重建图像均匀性。使用LV模型,一次填充锝,一次填充碘,通过对心肌活性和LV腔进行感兴趣区(ROI)分析,在重建图像上测量对比噪声比(CNR)。此外,进行极坐标图分析以确定水中溢出率(SOR)和图像噪声。将结果与双头平行孔低能高分辨率(LEHR)准直器系统的结果进行比较。对一名疑似冠状动脉疾病(CAD)的患者在LEHR系统上进行扫描,采用总采集时间为16分钟的局部协议,随后进行4分钟的MPH - 心脏型扫描。
在FOV的轴向中心发现峰值TS为1013 cps/MBq,而向径向边缘逐渐降低。对于LEHR和MPH - 心脏型,中心视野(CFOV)中的TS分别为134 cps/MBq和700 cps/MBq。MPH - 心脏型准直器在整个FOV的平均空间分辨率为4.38毫米半高宽(FWHM)。重建图像均匀性值在LEHR测量中为0.292%,在MPH - 心脏型测量中为0.214%。在锝的情况下,MPH - 心脏型的CNR高于LEHR(15.5对11.7),在碘的情况下也是如此(13.5对8.3)。对于LEHR和MPH - 心脏型,锝测量的SOR值分别为28.83%和21.1%,碘测量的SOR值分别为31.44%和24.33%。锝极坐标图的像素噪声值在LEHR中为0.38%,在MPH - 心脏型中为0.24%;碘极坐标图的像素噪声值在LEHR中为0.62%,在MPH - 心脏型中为0.21%。直观地解读患者扫描图像,与LEHR技术相比,MPH - 心脏型产生的图像对比度更好,扫描持续时间缩短了四倍。
在三头SPECT/CT系统上使用专用的MPH - 心脏型准直器实现的显著图像质量改善为短采集和低剂量心血管SPECT应用铺平了道路。