Sweeney R A
University of Missouri, Experiment Station Chemical Laboratories, Columbia, MO 65211.
J Assoc Off Anal Chem. 1989 Sep-Oct;72(5):770-4.
Nine laboratories participated in a collaborative study on determination of crude protein in animal feeds to compare a generically described combustion method with the AOAC mercury catalyst Kjeldahl method (7.015). The combustion method was written in general terms of method principle, apparatus specifications, and performance requirements. The sample set comprised closely matched pairs of feed ingredients and mixed products ranging from 10 to 90% protein. Ten pairs ground to 0.5 mm were the focus of the study; 4 pairs were ground to 1.0 mm for comparison. Nicotinic acid and lysine monohydrochloride were included as standards. Collaborators were instructed to report their results for performance checks using materials supplied. Only one laboratory failed to meet the proposed limits. Seven laboratories used the LECO Model FP-228 analyzer and 2 used the LECO CHN 600 analyzer. For the 0.5 mm pairs, repeatability standard deviations (Sr) ranged from 0.09 to 0.58 for the Kjeldahl method and from 0.14 to 0.33 for the combustion method, with a pooled Sr value of 0.28 and relative standard deviation (RSDr) of 0.59%. Reproducibility standard deviations (Sg) ranged from 0.23 to 0.86 (Kjeldahl) and from 0.30 to 0.61 (combustion), with a pooled Sg value of 0.52 and RSDg of 1.10%. Grand means for the samples ground to 0.5 mm were 47.65% protein by the combustion method and 47.41% protein by the Kjeldahl method. For samples ground to 1.0 mm, corresponding values were 31.82 and 31.50% protein. The generic combustion method has been approved interim official first action.
九个实验室参与了一项关于动物饲料中粗蛋白测定的协作研究,以比较一种通用描述的燃烧法与美国官方分析化学师协会(AOAC)的汞催化剂凯氏定氮法(7.015)。燃烧法以方法原理、仪器规格和性能要求的通用术语编写。样品集包括蛋白质含量从10%到90%的紧密匹配的饲料原料和混合产品对。研究重点是十对研磨至0.5毫米的样品;四对研磨至1.0毫米用于比较。烟酸和赖氨酸盐酸盐作为标准品包含在内。要求合作者使用提供的材料报告其性能检查结果。只有一个实验室未达到规定的限值。七个实验室使用LECO FP - 228型分析仪,两个实验室使用LECO CHN 600型分析仪。对于0.5毫米的样品对,凯氏定氮法的重复性标准偏差(Sr)范围为0.09至0.58,燃烧法的范围为0.14至0.33,合并的Sr值为0.28,相对标准偏差(RSDr)为0.59%。再现性标准偏差(Sg)范围为0.23至0.86(凯氏定氮法)和0.30至0.61(燃烧法),合并的Sg值为0.52,RSDg为1.10%。研磨至0.5毫米的样品的总平均值,燃烧法为蛋白质含量47.65%,凯氏定氮法为47.41%。对于研磨至1.0毫米的样品,相应的值分别为蛋白质含量31.82%和31.50%。通用燃烧法已被批准为暂行官方首次采用方法。