[智能手机检测喉镜灯光的准确性以及使用ISO标准对我们的喉镜进行审核]

[Accuracy of a smartphone to test laryngoscope's light and an audit to our laryngoscopes using an ISO standard].

作者信息

Machado Diogo Alcino de Abreu Ribeiro Carvalho, Esteves Dina da Assunção Azevedo, Branca Pedro Manuel Araújo de Sousa

机构信息

Hospital Pedro Hispano E.P.E., Unidade Local de Saúde de Matosinhos, Departamento de Anestesia, Matosinhos, Portugal.

Hospital Pedro Hispano E.P.E., Unidade Local de Saúde de Matosinhos, Bloco Operatório, Matosinhos, Portugal.

出版信息

Rev Bras Anestesiol. 2017 Mar-Apr;67(2):180-183. doi: 10.1016/j.bjan.2016.12.008. Epub 2017 Jan 23.

Abstract

BACKGROUND AND OBJECTIVES

Laryngoscope is a key tool in anesthetic practice. Direct laryngoscopy is a crucial moment and inadequate laryngoscope's light can lead to catastrophic consequences. From our experience laryngoscope's light is assessed in a subjective manner and we believe a more precise evaluation should be used. Our objective is to compare the accuracy of a smartphone compared to a lux meter. Secondly we audited our Operating Room laryngoscopes.

METHODS

We designed a pragmatic study, using as primary outcome the accuracy of a smartphone compared to the lux meter. Further we audited with both the lux meter and the smartphone all laryngoscopes and blades ready to use in our Operating Rooms, using the International Standard form the International Organization for Standardization.

RESULTS

For primary outcome we found no significant difference between devices. Our audit showed that only 2 in 48 laryngoscopes complied with the ISO norm. When comparing the measurements between the lux meter and the smartphone we found no significant difference.

DISCUSSION

Ideally every laryngoscope should perform as required. We believe all laryngoscopes should have a practical but reliable and objective test prior to its utilization. Our results suggest the smartphone was accurate enough to be used as a lux meter to test laryngoscope's light. Audit results showing only 4% comply with the ISO standard are consistent with other studies.

CONCLUSION

The tested smartphone has enough accuracy to perform light measurement in laryngoscopes. We believe this is a step further to perform an objective routine check to laryngoscope's light.

摘要

背景与目的

喉镜是麻醉实践中的关键工具。直接喉镜检查是一个关键时刻,喉镜光线不足可能导致灾难性后果。根据我们的经验,喉镜光线是通过主观方式评估的,我们认为应该采用更精确的评估方法。我们的目的是比较智能手机与照度计的准确性。其次,我们对手术室的喉镜进行了审核。

方法

我们设计了一项实用研究,将智能手机与照度计的准确性作为主要结果。此外,我们使用国际标准化组织的国际标准表格,用照度计和智能手机对手术室中所有准备使用的喉镜和镜片进行了审核。

结果

对于主要结果,我们发现不同设备之间没有显著差异。我们的审核表明,48台喉镜中只有2台符合ISO标准。在比较照度计和智能手机的测量结果时,我们没有发现显著差异。

讨论

理想情况下,每台喉镜都应按要求发挥作用。我们认为,所有喉镜在使用前都应进行实用但可靠且客观的测试。我们的结果表明,智能手机的准确性足以用作照度计来测试喉镜的光线。审核结果显示只有4%符合ISO标准,这与其他研究一致。

结论

经过测试的智能手机有足够的准确性来测量喉镜的光线。我们认为这是朝着对喉镜光线进行客观常规检查迈出的又一步。