Shenoy Niraj, Stenson Mary, Lawson Joshua, Abeykoon Jithma, Patnaik Mrinal, Wu Xiaosheng, Witzig Thomas
Division of Hematology, Mayo Clinic, Rochester, MN, USA.
Lab Invest. 2017 May;97(5):494-497. doi: 10.1038/labinvest.2017.18. Epub 2017 Feb 27.
Cell viability assays such as Cell Titer Blue and Alamar Blue rely on the reducing property of viable cells to reduce the reagent dye to a product which gives a fluorescent signal. The current manufacture-recommended protocols do not take into account the possibility of the reagent substrate being reduced directly to the fluorescent product by drugs with an anti-oxidant property. After suspecting spurious results while determining the cytotoxic potential of a drug of interest (DOI) with known anti-oxidant property against a renal cell cancer (RCC) cell line, we aimed to establish that drugs with anti-oxidant property can indeed cause false-negative results with the current protocols of these assays by direct reduction of the reagent substrate. We also aimed to counter the same with a simple modification added to the protocol. Through our experiments, we conclusively demonstrate that drugs with anti-oxidant properties can indeed interfere with cell viability measurements by assays that rely on the reducing property of viable cells. A simple modification in the protocol, as elaborated in the manuscript, can prevent spurious results with these otherwise convenient assays.Laboratory Investigation advance online publication, 27 February 2017; doi:10.1038/labinvest.2017.18.
诸如细胞活力蓝检测法和alamar蓝检测法等细胞活力检测方法,依赖活细胞的还原特性将试剂染料还原为能发出荧光信号的产物。目前生产厂家推荐的实验方案没有考虑到具有抗氧化特性的药物可能会直接将试剂底物还原为荧光产物的可能性。在怀疑一种已知具有抗氧化特性的目标药物(DOI)对肾癌细胞(RCC)细胞系的细胞毒性潜力检测结果存在虚假情况后,我们旨在证实具有抗氧化特性的药物确实会通过直接还原试剂底物,导致这些检测方法的现有实验方案出现假阴性结果。我们还旨在通过对实验方案进行简单修改来解决同样的问题。通过我们的实验,我们确凿地证明,具有抗氧化特性的药物确实会干扰依赖活细胞还原特性的检测方法对细胞活力的测量。正如手稿中所阐述的,对实验方案进行简单修改,可以防止这些原本便捷的检测方法出现虚假结果。《实验室研究》网络预发表,2017年2月27日;doi:10.1038/labinvest.2017.18