J Magn Reson Imaging. 2017 Dec;46(6):1673-1682. doi: 10.1002/jmri.25692. Epub 2017 Mar 16.
B mapping is important for many quantitative imaging protocols, particularly those that include whole-brain T mapping using the variable flip angle (VFA) technique. However, B mapping sequences are not typically available on many magnetic resonance imaging (MRI) scanners. The aim of this work was to demonstrate that B mapping implemented using standard scanner product pulse sequences can produce B (and VFA T ) maps comparable in quality and acquisition time to advanced techniques.
Six healthy subjects were scanned at 3.0T. An interleaved multislice spin-echo echo planar imaging double-angle (EPI-DA) B mapping protocol, using a standard product pulse sequence, was compared to two alternative methods (actual flip angle imaging, AFI, and Bloch-Siegert shift, BS). Single-slice spin-echo DA B maps were used as a reference for comparison (Ref. DA). VFA flip angles were scaled using each B map prior to fitting T ; the nominal flip angle case was also compared.
The pooled-subject voxelwise correlation (ρ) for B maps (BS/AFI/EPI-DA) relative to the reference B scan (Ref. DA) were ρ = 0.92/0.95/0.98. VFA T correlations using these maps were ρ = 0.86/0.88/0.96, much better than without B correction (ρ = 0.53). The relative error for each B map (BS/AFI/EPI-DA/Nominal) had 95 percentiles of 5/4/3/13%.
Our findings show that B mapping implemented using product pulse sequences can provide excellent quality B (and VFA T ) maps, comparable to other custom techniques. This fast whole-brain measurement (∼2 min) can serve as an excellent alternative for researchers without access to advanced B pulse sequences.
1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2017;46:1673-1682.
在许多定量成像协议中,B 映射都很重要,特别是那些包括使用可变翻转角(VFA)技术的全脑 T 映射的协议。然而,许多磁共振成像(MRI)扫描仪上通常没有 B 映射序列。本研究旨在证明,使用标准扫描仪产品脉冲序列实现的 B 映射可以生成质量和采集时间与先进技术相当的 B(和 VFA T)图。
在 3.0T 下对 6 名健康受试者进行扫描。使用标准产品脉冲序列的交错多切片自旋回波回波平面成像双角(EPI-DA)B 映射方案与两种替代方法(实际翻转角成像,AFI 和 Bloch-Siegert 位移,BS)进行比较。单切片自旋回波 DA B 图用作比较的参考(Ref. DA)。在拟合 T 之前,使用每个 B 图对 VFA 翻转角进行缩放;还比较了标称翻转角情况。
相对于参考 B 扫描(Ref. DA),B 图(BS/AFI/EPI-DA)的总体受试者体素相关性(ρ)为 ρ=0.92/0.95/0.98。使用这些图的 VFA T 相关性分别为 ρ=0.86/0.88/0.96,远优于没有 B 校正时的 ρ=0.53。每个 B 图(BS/AFI/EPI-DA/标称)的相对误差 95%置信区间为 5/4/3/13%。
我们的研究结果表明,使用产品脉冲序列实现的 B 映射可以提供高质量的 B(和 VFA T)图,与其他定制技术相当。这种快速的全脑测量(约 2 分钟)可以作为没有先进 B 脉冲序列的研究人员的极佳替代方案。