Suppr超能文献

单激发容积内插体部检查快速磁共振成像技术。

Reduced field of view single-shot spiral perfusion imaging.

机构信息

Departments of Medicine, Cardiovascular Division, University of Virginia Health System, Charlottesville, Virginia, USA.

Department of Radiology, Beth Israel Deaconess Medical Center, Boston, Massachusetts, USA.

出版信息

Magn Reson Med. 2018 Jan;79(1):208-216. doi: 10.1002/mrm.26664. Epub 2017 Mar 20.

Abstract

PURPOSE

To develop a single-shot spiral perfusion pulse sequence with outer-volume suppression (OVS) to achieve whole-heart coverage with a short temporal footprint of 10 ms per slice location.

METHODS

A highly accelerated single-shot variable density spiral pulse sequence with an integrated OVS module for reduced field of view (rFOV) perfusion imaging with 2 mm spatial resolution was developed and evaluated in simulations, phantom experiments and in clinical patients with (n = 8) or without (n = 8) OVS. Images were reconstructed by block low-rank sparsity with motion guidance (BLOSM) and graded by two cardiologists on a 5-point scale (1, excellent; 5, poor).

RESULTS

Simulation and phantom results showed that OVS effectively suppressed the signal outside the desired field of view (FOV). Clinical patient data demonstrated high quality perfusion images with rFOV. The average image quality scores of full FOV cases and rFOV cases were 3.1 ± 0.64 and 2.3 ± 0.46, respectively, (P = 0.02) from cardiologist 1 and 2.5 ± 0.54 and 1.8 ± 0.47, respectively, (P = 0.04) from cardiologist 2, showing superior image quality for the rFOV images compared with the full FOV images.

CONCLUSION

A single-shot spiral perfusion sequence that uses OVS and BLOSM performs perfusion imaging with a very short temporal footprint per image supporting whole-heart coverage with good image quality. Magn Reson Med 79:208-216, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

摘要

目的

开发一种具有外体积抑制(OVS)的单次螺旋灌注脉冲序列,以实现整个心脏覆盖,每个切片位置的时间足迹短至 10 毫秒。

方法

开发了一种高度加速的单次变量密度螺旋脉冲序列,该序列具有集成的 OVS 模块,用于具有 2mm 空间分辨率的小视野(rFOV)灌注成像,并在模拟、体模实验和具有(n=8)或不具有(n=8)OVS 的临床患者中进行了评估。图像通过块低秩稀疏性与运动引导(BLOSM)重建,并由两位心脏病专家进行 5 分制评分(1,优秀;5,差)。

结果

模拟和体模结果表明,OVS 有效地抑制了期望视野(FOV)之外的信号。临床患者数据显示 rFOV 具有高质量的灌注图像。来自心脏病专家 1 的全 FOV 病例和 rFOV 病例的平均图像质量评分分别为 3.1±0.64 和 2.3±0.46(P=0.02),来自心脏病专家 2 的全 FOV 病例和 rFOV 病例的平均图像质量评分分别为 2.5±0.54 和 1.8±0.47(P=0.04),rFOV 图像的图像质量优于全 FOV 图像。

结论

一种使用 OVS 和 BLOSM 的单次螺旋灌注序列可在每张图像上进行非常短的时间足迹灌注成像,支持整个心脏覆盖,并具有良好的图像质量。磁共振医学 79:208-216,2018。©2017 国际磁共振学会。