Gould B A, Turner J, Keeling D H, Ring N J, Cox R R, Marshall A J
Department of Cardiology, Plymouth Group of Hospitals.
Br Heart J. 1988 Apr;59(4):463-7. doi: 10.1136/hrt.59.4.463.
A cadmium telluride nuclear probe with an Elscint gamma camera was used to detect and measure left to right intracardiac shunts at the bedside in 34 patients. Fifteen also had right heart catheterisation and oximetric measurement of the shunt. For the nuclear technique 740 MBq (20 mCi) of technetium-99m pertechnetate was injected into the right antecubital vein and the pulmonary to systemic flow ratio (QP:QS) was measured by the gamma variate technique. Data were not obtained in four patients because the nuclear probe failed in three patients and one storage disc was corrupted. Data from the gamma camera were lost in another patient. When the size of the shunt measured by the nuclear probe was compared with that measured by the oximetric technique the mean difference (SD of mean difference) was 0.36 (SD 0.78) and when it was compared with the gamma camera it was 0.08 (SD 0.67). Analysis of scatter plots showed that the larger the shunt, the larger the discrepancy. Twenty four of 29 data sets showed complete agreement between the nuclear probe and gamma camera on the size of the shunt. Any differences were small. These data indicate that left to right intracardiac shunts may be measured accurately by a nuclear probe at the bedside in either the coronary care unit or outpatient department.
采用配备Elscintγ相机的碲化镉核探头,在床边对34例患者的心内左向右分流进行检测和测量。其中15例患者还接受了右心导管检查及分流的血氧测定。对于核技术,将740 MBq(20 mCi)的99m锝高锝酸盐注入右肘前静脉,并采用γ变量技术测量肺循环与体循环血流量之比(QP:QS)。4例患者未获得数据,原因是3例患者的核探头出现故障,1个存储盘损坏。另1例患者γ相机的数据丢失。将核探头测量的分流大小与血氧测定技术测量的结果进行比较,平均差异(平均差异标准差)为0.36(标准差0.78),与γ相机测量结果比较时平均差异为0.08(标准差0.67)。散点图分析显示,分流越大,差异越大。29个数据集中有24个显示核探头与γ相机在分流大小方面完全一致。任何差异都很小。这些数据表明,在冠心病监护病房或门诊,可使用核探头在床边准确测量心内左向右分流。