Magnetization and demagnetization of magnetic dental attachments in a 3-T MRI system.

作者信息

Hayashi Norio, Ogura Akio, Tsuchihashi Toshio, Takahashi Daisuke, Matsuda Tsuyoshi, Seino Shinya, Doi Tsukasa

机构信息

Department of Radiological Technology, Gunma Prefectural College of Health Sciences, 323-1 Kamioki, Maebashi, Gunma, 371-0052, Japan.

Department of Radiology, Nippon Medical School Hospital, Tokyo, Japan.

出版信息

Radiol Phys Technol. 2017 Sep;10(3):294-300. doi: 10.1007/s12194-017-0399-0. Epub 2017 Apr 27.

Abstract

The purpose of this study was to evaluate the magnetization and demagnetization of magnetic dental attachments in a 3.0-T magnetic resonance imaging (MRI) scanner. A high-field-strength (3.0 T) MRI scanner (Achieva 3.0 T, Philips, Amsterdam, Netherlands) was used. Magnetic flux leakage was measured using a gaussmeter. To evaluate the magnetization and demagnetization of the magnetic assemblies and keepers caused by the moving in and out of the MRI scanner, the magnetic force of the attachments was measured before and after the table was moved in and out. Two settings were used wherein the magnetic assemblies and keepers were positioned on the table at angles of 0° or 90° with respect to the magnetic flux of the static magnetic field. The movement of the table was repeated 15 times. In addition, the retentive force of the magnetic dental attachments was measured after magnetic field exposure. The magnetic force of the magnetic attachments positioned at 0° decreased significantly after moving in and out of the MRI scanner (p < 0.05). In contrast, the magnetic force of the magnetic attachments positioned at 90° was stable after the movement of the table. The magnetic force of the keepers placed at both 0° and 90° was slightly increased after the movement of the table. At 0°, the retentive force of the magnet-keeper combinations decreased when the magnetic assembly was exposed to the strong magnetic flux of the MRI scanner. Therefore, the removal of all removable magnetic dentures is recommended before an MRI examination.

摘要

本研究的目的是评估磁性牙科附件在3.0-T磁共振成像(MRI)扫描仪中的磁化和退磁情况。使用了一台高场强(3.0 T)MRI扫描仪(Achieva 3.0 T,飞利浦,荷兰阿姆斯特丹)。使用高斯计测量磁通量泄漏。为了评估由MRI扫描仪的进出移动引起的磁性组件和保持器的磁化和退磁情况,在工作台进出之前和之后测量附件的磁力。使用了两种设置,其中磁性组件和保持器相对于静磁场的磁通量以0°或90°的角度放置在工作台上。工作台的移动重复15次。此外,在磁场暴露后测量磁性牙科附件的保持力。进出MRI扫描仪后,位于0°的磁性附件的磁力显著降低(p < 0.05)。相比之下,工作台移动后,位于90°的磁性附件的磁力保持稳定。工作台移动后,放置在0°和90°的保持器的磁力略有增加。在0°时,当磁性组件暴露于MRI扫描仪的强磁通量时,磁体-保持器组合的保持力降低。因此,建议在MRI检查前取下所有可摘磁性假牙。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验