Department of Chemistry, Purdue University , 560 Oval Drive, West Lafayette, Indiana 47907, United States.
Department of Attribute Sciences, Amgen Inc. , 1 Amgen Center Drive, Thousand Oaks, California 91320, United States.
Anal Chem. 2017 Jul 5;89(13):6969-6975. doi: 10.1021/acs.analchem.7b00119. Epub 2017 Jun 15.
Advances in chemical sampling using miniature mass spectrometer technology are used to monitor slow reactions at a frequency of ca. 180 h (on the Mini 12) with no sample carryover and with inline derivatization in the case of poorly ionizing compounds. Moreover, we demonstrate high reproducibility with a relative error of less than 10% for major components. Monitoring is enabled using a continuous-flow nanoelectrospray (CF-nESI) probe contained in a custom-built 3D-printed rotary holder. The holder position is automatically set using a stepper motor controlled by a microcontroller. Reaction progress of up to six reactions, including hydrazone formation and Katritzky transamination, can be monitored simultaneously without carryover for several hours.
微型质谱仪技术在化学采样方面的进展用于以约 180 h 的频率(在 Mini 12 上)监测缓慢反应,且不存在样品残留,并可对电离性能差的化合物进行在线衍生化。此外,我们通过相对误差小于 10%的方式证明了主要成分的高重现性。监测是使用包含在定制 3D 打印旋转支架中的连续流纳喷雾(CF-nESI)探头实现的。支架位置使用由微控制器控制的步进电机自动设置。可以同时监测多达六个反应的反应进度,包括腙形成和 Katritzky 转氨反应,且在数小时内不存在样品残留。