Accurate determination of brain metabolite concentrations using ERETIC as external reference.

作者信息

Zoelch Niklaus, Hock Andreas, Heinzer-Schweizer Susanne, Avdievitch Nikolai, Henning Anke

机构信息

Institute for Biomedical Engineering, University and ETH  Zurich, Zurich, Switzerland.

Institute of Forensic Medicine, University of Zurich, Zurich, Switzerland.

出版信息

NMR Biomed. 2017 Aug;30(8). doi: 10.1002/nbm.3731. Epub 2017 May 25.

Abstract

Magnetic Resonance Spectroscopy (MRS) can provide in vivo metabolite concentrations in standard concentration units if a reliable reference signal is available. For H MRS in the human brain, typically the signal from the tissue water is used as the (internal) reference signal. However, a concentration determination based on the tissue water signal most often requires a reliable estimate of the water concentration present in the investigated tissue. Especially in clinically interesting cases, this estimation might be difficult. To avoid assumptions about the water in the investigated tissue, the Electric REference To access In vivo Concentrations (ERETIC) method has been proposed. In this approach, the metabolite signal is compared with a reference signal acquired in a phantom and potential coil-loading differences are corrected using a synthetic reference signal. The aim of this study, conducted with a transceiver quadrature head coil, was to increase the accuracy of the ERETIC method by correcting the influence of spatial B inhomogeneities and to simplify the quantification with ERETIC by incorporating an automatic phase correction for the ERETIC signal. Transmit field ( B1+) differences are minimized with a volume-selective power optimization, whereas reception sensitivity changes are corrected using contrast-minimized images of the brain and by adapting the voxel location in the phantom measurement closely to the position measured in vivo. By applying the proposed B correction scheme, the mean metabolite concentrations determined with ERETIC in 21 healthy subjects at three different positions agree with concentrations derived with the tissue water signal as reference. In addition, brain water concentrations determined with ERETIC were in agreement with estimations derived using tissue segmentation and literature values for relative water densities. Based on the results, the ERETIC method presented here is a valid tool to derive in vivo metabolite concentration, with potential advantages compared with internal water referencing in diseased tissue.

摘要

如果有可靠的参考信号,磁共振波谱(MRS)能够以标准浓度单位提供体内代谢物浓度。对于人脑的氢质子MRS,通常将组织水的信号用作(内部)参考信号。然而,基于组织水信号的浓度测定通常需要对被研究组织中存在的水浓度进行可靠估计。特别是在临床相关的情况下,这种估计可能很困难。为了避免对被研究组织中的水进行假设,人们提出了用于获取体内浓度的电参考(ERETIC)方法。在这种方法中,将代谢物信号与在体模中获取的参考信号进行比较,并使用合成参考信号校正潜在的线圈负载差异。本研究使用收发一体式正交头线圈进行,目的是通过校正空间磁场不均匀性的影响来提高ERETIC方法的准确性,并通过对ERETIC信号进行自动相位校正来简化ERETIC定量分析。通过体积选择性功率优化将发射场(B1 +)差异降至最低,而接收灵敏度变化则通过大脑的对比度最小化图像以及使体模测量中的体素位置紧密适应体内测量位置来校正。通过应用所提出的磁场校正方案,在21名健康受试者的三个不同位置用ERETIC测定的平均代谢物浓度与以组织水信号为参考得出的浓度一致。此外用ERETIC测定的脑水浓度与使用组织分割和相对水密度文献值得出的估计值一致。基于这些结果,本文介绍的ERETIC方法是得出体内代谢物浓度的有效工具,与在患病组织中使用内部水参考相比具有潜在优势。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验