Agreement of invasive and non-invasive blood pressure measurements in anaesthetised pigs using the Surgivet V9203.

作者信息

Tuohy Philomena P, Raisis Anthea L, Drynan Eleanor A

机构信息

College of Veterinary Medicine, School of Veterinary and Life Sciences, Murdoch University, Western Australia 6150, Australia.

出版信息

Res Vet Sci. 2017 Dec;115:250-254. doi: 10.1016/j.rvsc.2017.05.022. Epub 2017 May 19.

Abstract

The automated oscillometric method is a common method for measuring blood pressure non-invasively and is broadly and confidently used in the veterinary setting. Twenty-one pigs undergoing anesthesia for exploratory laparotomy were enrolled in a study to evaluate the performance of the Surgivet V9203 non-invasive blood pressure (NIBP) monitor. The aim was to compare measurements of arterial blood pressure obtained simultaneously using the Surgivet V9203 oscillometric system and an intra-arterial catheter-transducer system, both at the level of the metatarsus. Invasive blood pressure (IBP) was consistently underestimated by the non-invasive method. Bland-Altman analysis showed poor agreement between the two modalities based on evaluation of mean bias and 95% limits of agreement. The Surgivet V9203 cannot therefore be recommended as a reliable alternative to invasive blood pressure monitoring in anaesthetised pigs. As pulse detection is one of the most important factors affecting NIBP accuracy it is likely that our findings may reflect an anatomical or physiological difference in the species that alters the detection of wall movement by the oscillometric technique and additionally, makes the algorithm used by the Surgivet to determine blood pressure parameters unsuitable for use in pigs.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验