Vu Anh Phuong, Nguyen Thi Ngan, Do Thi Trang, Doan Thu Ha, Ha Tran Hung, Ta Thi Thao, Nguyen Hung Long, Hauser Peter C, Nguyen Thi Anh Huong, Mai Thanh Duc
Poison Control Center, Bach Mai Hospital, 78 Giai Phong road, Dong Da, Hanoi, Viet Nam; Department of Analytical Chemistry, Faculty of Chemistry, VNU University of Science, Vietnam National University, Hanoi - 19 Le Thanh Tong, Hanoi, Viet Nam(1).
Poison Control Center, Bach Mai Hospital, 78 Giai Phong road, Dong Da, Hanoi, Viet Nam.
J Chromatogr B Analyt Technol Biomed Life Sci. 2017 Aug 15;1060:111-117. doi: 10.1016/j.jchromb.2017.06.010. Epub 2017 Jun 12.
The employment of a purpose-made capillary electrophoresis (CE) instrument with capacitively coupled contactless conductivity detection (CD) as a simple and cost-effective solution for clinical screening of paraquat in plasma samples for early-stage diagnosis of acute herbicide poisoning is reported. Paraquat was determined using an electrolyte composed of 10mM histidine adjusted to pH 4 with acetic acid. A detection limit of 0.5mg/L was achieved. Good agreement between results from CE-CD and the confirmation method (HPLC-UV) was obtained, with relative errors for the two pairs of data better than 20% for 31 samples taken from paraquat-intoxicated patients. The results were used by medical doctors for identification and prognosis of acute paraquat poisoning cases. The objective of the work is the deployment of the developed approach in rural areas in Vietnam as a low-cost solution to reduce the mortality rate due to accidental or suicidal ingestion of paraquat.
报道了使用配备电容耦合非接触式电导检测(CD)的专用毛细管电泳(CE)仪器,作为一种简单且经济高效的解决方案,用于临床筛查血浆样本中的百草枯,以实现急性除草剂中毒的早期诊断。使用由10mM组氨酸组成、用乙酸调节至pH 4的电解质来测定百草枯。实现了0.5mg/L的检测限。CE-CD结果与确认方法(HPLC-UV)之间取得了良好的一致性,对于从百草枯中毒患者采集的31个样本,两对数据的相对误差优于20%。这些结果被医生用于急性百草枯中毒病例的识别和预后判断。这项工作的目标是在越南农村地区采用所开发的方法,作为一种低成本解决方案,以降低因意外或自杀性摄入百草枯导致的死亡率。