Suppr超能文献

使用压缩感知的三维黑血多对比颈动脉成像:一项重复性研究。

Three-dimensional black-blood multi-contrast carotid imaging using compressed sensing: a repeatability study.

作者信息

Yuan Jianmin, Usman Ammara, Reid Scott A, King Kevin F, Patterson Andrew J, Gillard Jonathan H, Graves Martin J

机构信息

Department of Radiology, School of Clinical Medicine, University of Cambridge, Level 5, Box 218, Addenbrooke's Hospital, Hills Rd, Cambridge, CB2 0QQ, UK.

GE Healthcare, Amersham, UK.

出版信息

MAGMA. 2018 Feb;31(1):183-190. doi: 10.1007/s10334-017-0640-1. Epub 2017 Jun 26.

Abstract

OBJECTIVE

The purpose of this work is to evaluate the repeatability of a compressed sensing (CS) accelerated multi-contrast carotid protocol at 3 T.

MATERIALS AND METHODS

Twelve volunteers and eight patients with carotid disease were scanned on a 3 T MRI scanner using a CS accelerated 3-D black-blood multi-contrast protocol which comprises T w, T w and PDw without CS, and with a CS factor of 1.5 and 2.0. The volunteers were scanned twice, the lumen/wall area and wall thickness were measured for each scan. Eight patients were scanned once, the inter/intra-observer reproducibility of the measurements was calculated.

RESULTS

In the repeated volunteer scans, the interclass correlation coefficient (ICC) for the wall area measurement using a CS factor of 1.5 in PDw, T w and T w were 0.95, 0.81, and 0.97, respectively. The ICC for lumen area measurement using a CS factor of 1.5 in PDw, T w and T w were 0.96, 0.92, and 0.96, respectively. In patients, the ICC for inter/intra-observer measurements of lumen/wall area, and wall thickness were all above 0.81 in all sequences.

CONCLUSION

The results show a CS accelerated 3-D black-blood multi-contrast protocol is a robust and reproducible method for carotid imaging. Future protocol design could use CS to reduce the scanning time.

摘要

目的

本研究旨在评估3T磁共振成像中压缩感知(CS)加速多对比颈动脉成像方案的可重复性。

材料与方法

12名志愿者和8名患有颈动脉疾病的患者在3T磁共振成像扫描仪上采用CS加速三维黑血多对比成像方案进行扫描,该方案包括无CS时的T1加权、T2加权和质子密度加权成像,以及CS因子分别为1.5和2.0时的成像。志愿者进行了两次扫描,每次扫描测量管腔/管壁面积和管壁厚度。8名患者进行了一次扫描,计算测量的观察者间/观察者内重复性。

结果

在志愿者的重复扫描中,PDw、T1加权和T2加权成像中CS因子为1.5时管壁面积测量的组内相关系数(ICC)分别为0.95、0.81和0.97。PDw、T1加权和T2加权成像中CS因子为1.5时管腔面积测量的ICC分别为0.96、0.92和0.96。在患者中,所有序列中观察者间/观察者内测量的管腔/管壁面积和管壁厚度的ICC均高于0.81。

结论

结果表明,CS加速三维黑血多对比成像方案是一种可靠且可重复的颈动脉成像方法。未来的成像方案设计可采用CS来缩短扫描时间。