To compare objective image quality at distant, intermediate, and near foci with the corresponding visual acuity (VA) in patients symmetrically implanted with 5 different diffractive multifocal intraocular lenses (IOLs) (ReSTOR SV25T0, Tecnis ZKB00, Tecnis ZLB00, AT LISA 809, and AT LISA Tri 839MP) and a monofocal lens (Tecnis ZA9003) 3 months after cataract intervention.
Objective image quality, measured as the area under the modulation transfer function curve (AMTF), was tested in vitro in an eye model. In addition, corrected distance visual acuity and distance-corrected intermediate and near visual acuities (DCIVA and DCNVA) were assessed in a group of 79 patients, randomly implanted with one of the lenses.
Image quality and VA at the distant focus was good for all lens designs, but was similarly compromised at the intermediate focus, where the best DCIVA corresponded to the low add ZKB00. At the near focus, the monofocal lens and the distance dominant SV25T0 had the worst AMTF values, with the ZLB00 presenting the best DCNVA. The relationship between AMTF and VA was found not to be linear, with VA being similarly good for AMTF values over a certain threshold.
Visual acuity of pseudophakic patients reflected the optical quality of the specific IOL design within a certain range of measurements, beyond which maximum VA is limited by other ocular, optical, and neuropsychophysical factors.
比较在白内障干预3个月后,对称植入5种不同衍射多焦点人工晶状体(IOL)(ReSTOR SV25T0、Tecnis ZKB00、Tecnis ZLB00、AT LISA 809和AT LISA Tri 839MP)和一种单焦点晶状体(Tecnis ZA9003)的患者在远、中、近焦点处的客观图像质量与相应的视力(VA)。
在体外眼模型中测试以调制传递函数曲线下面积(AMTF)衡量的客观图像质量。此外,对一组79例随机植入其中一种晶状体的患者评估矫正远视力以及距离矫正后的中视力和近视力(DCIVA和DCNVA)。
所有晶状体设计在远焦点处的图像质量和视力均良好,但在中焦点处同样受到损害,其中最佳DCIVA对应低附加度数的ZKB00。在近焦点处,单焦点晶状体和距离主导型的SV25T0的AMTF值最差,而ZLB00的DCNVA最佳。发现AMTF与VA之间的关系并非线性,在一定阈值以上的AMTF值对应的VA同样良好。
人工晶状体眼患者视力在一定测量范围内反映了特定IOL设计的光学质量,超过该范围,最大视力受其他眼部、光学和神经心理物理学因素限制。