Abstract

INTRODUCTION

In this study, analytic performance (imprecision, carryover, time stability) and diagnostic efficiency of Mindray BC-6800 analyzer to quantify reticulocytes and extended reticulocyte parameters was evaluated. Moreover, reference intervals on adult population were determined. Results were compared with those obtained by Sysmex XE-5000 analyzer.

METHODS

One hundred and eighty-four healthy adults of both sexes, and 368 subjects affected by various pathologic conditions (nutritional anemias before and after treatment, hemolytic and posthemorragic anemias, acute and chronic inflammations, malignancy under therapy, and beta thalassemia trait) were selected.

RESULTS

Reference intervals were as follows: reticulocytes (×10 /L): 23.2-93.2; immature reticulocyte fraction: 0.015-0.14; mean reticulocyte hemoglobin equivalent (RHE) (pg): 30.9-35.7; mean reticulocyte volume (fL): 97.8-118. Imprecision on reticulocyte count at all concentrations was close to analytic goal based on within-subject biological variation. Carryover (2.3%) was negligible, and time-stability was excellent up to 8 hours.

CONCLUSION

When compared with XE-5000, BC-6800 shows a good overall correlation on counting despite evidence of difference in the upper limit of reference intervals (93.2 vs 101.3). Comparison of diagnostic efficiency of extended parameters shows a good total agreement of RHE (91.6%).

摘要

引言

在本研究中,评估了迈瑞BC - 6800分析仪定量网织红细胞及扩展网织红细胞参数的分析性能(不精密度、携带污染、时间稳定性)和诊断效率。此外,还确定了成人人群的参考区间。将结果与Sysmex XE - 5000分析仪获得的结果进行比较。

方法

选取184名健康成年男女以及368名患有各种病理状况的受试者(治疗前后的营养性贫血、溶血性和失血性贫血、急慢性炎症、接受治疗的恶性肿瘤以及β地中海贫血特征)。

结果

参考区间如下:网织红细胞(×10⁹/L):23.2 - 93.2;未成熟网织红细胞比例:0.015 - 0.14;平均网织红细胞血红蛋白含量(RHE)(pg):30.9 - 35.7;平均网织红细胞体积(fL):97.8 - 118。基于受试者内生物学变异,所有浓度下网织红细胞计数的不精密度接近分析目标。携带污染率(2.3%)可忽略不计,且长达8小时的时间稳定性极佳。

结论

与XE - 5000相比,尽管参考区间上限存在差异(93.2对101.3),但BC - 6800在计数方面显示出良好的总体相关性。扩展参数诊断效率的比较显示RHE的总体一致性良好(91.6%)。

相似文献

8
Automated reticulocyte counting using the Sysmex RAM-1.使用Sysmex RAM-1进行自动网织红细胞计数。
Clin Lab Haematol. 2001 Apr;23(2):97-102. doi: 10.1046/j.1365-2257.2001.00365.x.
10
Clinical utility of reticulocyte parameters.网织红细胞参数的临床应用价值
Clin Lab Med. 2015 Mar;35(1):133-63. doi: 10.1016/j.cll.2014.10.004. Epub 2014 Nov 26.

引用本文的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验