Lee Hyung Jin, Yoon Dae Young, Seo Young Lan, Kim Jin Ho, Baek Sora, Lim Kyoung Ja, Cho Young Kwon, Yun Eun Joo
Department of Radiology, Kangdong Seong-Sim Hospital, Hallym University College of Medicine, Seoul, Korea.
J Ultrasound Med. 2018 Jan;37(1):173-178. doi: 10.1002/jum.14316. Epub 2017 Jul 24.
The purpose of this study was to assess the intraobserver and interobserver variability in ultrasound (US) measurements of thyroid nodules.
We performed a prospective study of the US examinations of 73 patients with 122 thyroid nodules greater than 5 mm in size. Ultrasound measurements in 4 dimensions (anteroposterior, transverse, longitudinal, and maximum diameters) and measurement of the estimated volume (using the ellipsoid formula) of each thyroid nodule were performed twice by 2 independent radiologists (A and B, with 10 years and 6 months of experience, respectively). The intraobserver and interobserver variability in measurements of thyroid nodules was assessed by a Bland-Altman analysis of agreement. The absolute values for intraobserver and interobserver variability were compared by a paired t test.
The 95% intraobserver and interobserver limits of agreement for the anteroposterior, transverse, longitudinal, and maximum diameters and estimated volume of thyroid nodules were ±18.2%, ± 14.3%, and ±21.0%; ± 17.2%%, ± 17.3%, and 18.2%; ± 14.6%, ± 15.5%, and ±22.3%; ± 13.8%, ± 15.5%, and ±19.6%; and ±30.2%, ± 27.7%, and ±44.1%, respectively. The absolute values for intraobserver variability were lower than those for interobserver variability for all measurements.
There was considerable intraobserver and interobserver variability in US measurement of thyroid nodules, which must be taken into account during follow-up US examinations of patients with thyroid nodules.
本研究旨在评估甲状腺结节超声(US)测量中的观察者内和观察者间变异性。
我们对73例患者的122个大小大于5毫米的甲状腺结节进行了US检查的前瞻性研究。两名独立的放射科医生(A和B,分别有10年和6个月的经验)对每个甲状腺结节的4个维度(前后径、横径、纵径和最大直径)进行超声测量,并测量估计体积(使用椭球体公式),各进行两次。通过Bland-Altman一致性分析评估甲状腺结节测量中的观察者内和观察者间变异性。通过配对t检验比较观察者内和观察者间变异性的绝对值。
甲状腺结节前后径、横径、纵径、最大直径和估计体积的观察者内和观察者间95%一致性界限分别为±18.2%、±14.3%和±21.0%;±17.2%、±17.3%和18.2%;±14.6%、±15.5%和±22.3%;±13.8%、±15.5%和±19.6%;以及±30.2%、±27.7%和±44.1%。所有测量中观察者内变异性的绝对值均低于观察者间变异性的绝对值。
甲状腺结节US测量中存在相当大的观察者内和观察者间变异性,在对甲状腺结节患者进行随访US检查时必须予以考虑。