Nowak Roxana, Wesemann Christian, Robben Jan, Muallah Jonas, Bumann Axel
Quintessence Int. 2017 Jul 20:667-676. doi: 10.3290/j.qi.a38676.
The purpose of this study was to evaluate the accuracy of full-arch scans using 11 different cast scanners.
Dental arch length (AL), intermolar width (IMW), and intercanine width (ICW) of the reference plaster casts were measured using a coordinate measuring machine (Zeiss O-Inspect 422). The master cast was subsequently scanned 37 times using 11 desktop scanners (3Shape R500, 3Shape R700, 3Shape R1000, 3Shape R2000, Medianetx grande, Medianetx colori, DentaCore CS ULTRA, Dentaurum OrthoX, Maestro 3D, Imetric IScan D104i, GC Aadva Lab Scan). Using the software Convince Premium 2012 (3Shape), AL, IMW, and ICW were measured on the digital models and compared to the reference plaster cast.
The accuracy of the cast scanners differed significantly. The most accurate measurements were given by the cast scanners 3Shape R700 (ICW: 7.4?±?5.9?µm) and Imetric IScan D104i (ICW: 9.1?±?4.9?µm). The cast scanners 3Shape R1000 (ICW: 11.2?±?3.4?µm) and GC Aadva Lab Scan (ICW: 13.8?±?8.1?µm) yielded comparable measurements. DentaCore (ICW: 26.6?±?7.5?µm) and Dentaurum OrthoX (ICW: 31.1?±?24?µm) were significantly less accurate.
Almost all scanners demonstrated a level of accuracy so high that further improvement would not provide additional benefit for use in orthodontics. Advancement of the scanners should focus primarily on reducing time and cost. For prosthodontic use, the scanners with the highest accuracy are recommended.
本研究旨在评估使用11种不同的石膏模型扫描仪进行全牙弓扫描的准确性。
使用坐标测量机(蔡司O-Inspect 422)测量参考石膏模型的牙弓长度(AL)、磨牙间宽度(IMW)和尖牙间宽度(ICW)。随后使用11台台式扫描仪(3Shape R500、3Shape R700、3Shape R1000、3Shape R2000、Medianetx grande、Medianetx colori、DentaCore CS ULTRA、Dentaurum OrthoX、Maestro 3D、Imetric IScan D104i、GC Aadva Lab Scan)对主模型进行37次扫描。使用Convince Premium 2012软件(3Shape)在数字模型上测量AL、IMW和ICW,并与参考石膏模型进行比较。
石膏模型扫描仪的准确性差异显著。最准确的测量结果由3Shape R700石膏模型扫描仪(ICW:7.4±5.9μm)和Imetric IScan D104i(ICW:9.1±4.9μm)给出。3Shape R1000石膏模型扫描仪(ICW:11.2±3.4μm)和GC Aadva Lab Scan(ICW:13.8±8.1μm)得出的测量结果相当。DentaCore(ICW:26.6±7.5μm)和Dentaurum OrthoX(ICW:31.1±24μm)的准确性明显较低。
几乎所有扫描仪都显示出非常高的准确性水平,进一步改进对正畸应用不会带来额外益处。扫描仪的改进应主要集中在减少时间和成本上。对于修复应用,建议使用准确性最高的扫描仪。